English
Language : 

81250 Datasheet, PDF (55/64 Pages) Keysight Technologies – Agilent ParBERT 81250 Parallel Bit Error Ratio Tester
General characteristics
Mainframes: See table 72 on
page 56.
Save/recall: Pattern segments,
settings and complete settings
plus segments can be saved and
recalled. The number of settings
that can be stored is limited only
by disk space.
Vector import/export: File format is
ASCII using a STIL subset.
Import/export to/from ParBERT
internal data-base for regular file
transfer via USB memory stick, CD
or LAN is handled by ParBERT
user SW. Direct programming of
data segments is possible via
GPIB (IEEE 488.2) and SCPI.
Programming interface: GP-IB
(IEEE 488.2) and LAN. The
interface to applications such as
C, Visual Basic, or VEE must be
installed. Use the Agilent 81200
Plug&Play drivers for easy
programming.
Programming language:
SCPI 1992.0, active X for MUI
Programming times: Vector transfer
from memory to hardware
depends on the amount of data.
See Table 70 for examples.
On-line help: Context-sensitive.
Print-on-demand: Getting started
and programming guides can be
printed from .pdf files included
in the ParBERT 81250 software.
Self-test: Module and system
selftests can be initiated.
Modules
Module size:
VXI C-size, 1 slot.
Module type:
Register-based; requires
ParBERT 81250 user software
E4875A supplied with the
mainframes.
Weight:
(including front-ends) Net: 2kg.
Shipping: 2.5 kg.
Warranty:
1 year return to Agilent
Re-calibration period: 1 year.
Agilent Technologies
quality standards
The ParBERT 81250 is produced
in accordance to the ISO 9001
international quality system
standard as part of Agilent
Technologies’ commitment to
continually increasing customer
satisfaction through improved
quality control.
Table 70. Programming times
Parameter
Change of levels
Change of delay
Change of period (1)
Stop & start
Synchronization (1)
Download values
System with 4 channels
System with 120 channels
System with 40 channels
Programming time
For one E4805B with one E4832A.
Increases with the number of modules.
6 ms typ.
16 ms typ. Not applicable in run mode
60 ms typ. Not applicable in run mode.
32 ms typ.
50 ms typ. (without phase alignment)
110 ms typ. with 20% phase accuracy @ 660 MHz
650 ms typ. with 1% phase accuracy @ 660 MHz
< 1.5 s typ. 1000,000 bit each
< 20 s typ. 1 Mbit each
< 10 s typ. 1 Mbit each
(1) valid for a system consisting of one E4805A and one E4832A.
Table 71. Cooling requirements for modules with front-ends installed
Modules
E4805B
E4808A
E4809A
E4832A
E4861B
N4872A/74A
N4873A/75A
∆P mm H2O
0.25
0.25
1.08
0.30
0.40
1.20
0.85
Air flow liter/s
3.6
3.6
3.7
4.7
6.6
7.5
5.4
Max ∆temp
10 °C
10 °C
12 °C
15 °C
15 °C
13 °C
12 °C
ParBERT 81250 Main Overview