81250 Datasheet, PDF (12/64 Pages) Keysight Technologies – Agilent ParBERT 81250 Parallel Bit Error Ratio Tester | |||
|
◁ |
Eye opening
During this measurement both
sampling point timing and
analyzer threshold voltage are
automatically swept, while BER
is recorded and plotted either as
a pseudo-color- or a contour-plot
(see figure 9 a,b,c), giving the
user intuitive and in-depth
information of the pulse- or
eye-performance of the RX
under test.
Range and resolution for the
measurement parameters and
pass/fail values are user
definable.
Table 4: Eye opening
Numerical Measurement Results
Pass/fail
Graphical Result Displays
Optimum sample Point delay
Optimum threshold
Eye opening (Volt)Phase margin
For all parameters
Each parameter can be individually enabled
pseudo color plot BER-contour
Two markers: voltage, delay, BER
Figure 9a/b/c. View the BER for one terminal as a pseudo color plot or contour plot or equal BER at BER threshold
Page 12/64
ParBERT 8125012Main Overview
|
▷ |