Register description
LSM6DS3H
9.75
WAKE_UP_THS (5Bh)
Single and double-tap function threshold register (r/w).
Table 184. WAKE_UP_THS register
SINGLE_
DOUBLE INACTIVITY WK_THS5 WK_THS4 WK_THS3 WK_THS2 WK_THS1 WK_THS0
_TAP
SINGLE_
DOUBLE_TAP
INACTIVITY
WK_THS[5:0]
Table 185. WAKE_UP_THS register description
Single/double-tap event enable. Default: 0
(0: only single-tap event enabled; 1: both single and double-tap events enabled)
Inactivity event enable. Default value: 0
(0: sleep disabled; 1: sleep enabled)
Threshold for wakeup. Default value: 000000
9.76
WAKE_UP_DUR (5Ch)
Free-fall, wakeup, timestamp and sleep mode functions duration setting register (r/w).
FF_DUR5
WAKE_
DUR1
Table 186. WAKE_UP_DUR register
WAKE_ TIMER_ SLEEP_ SLEEP_
DUR0
HR
DUR3
DUR2
SLEEP_
DUR1
SLEEP_
DUR0
FF_DUR5
Table 187. WAKE_UP_DUR register description
Free fall duration event. Default: 0
For the complete configuration of the free-fall duration, refer to FF_DUR[4:0] in
FREE_FALL (5Dh) configuration.
WAKE_DUR[1:0]
TIMER_HR
Wake up duration event. Default: 00
1LSB = 1 ODR_time
Timestamp register resolution setting(1). Default value: 0
(0: 1LSB = 6.4 ms; 1: 1LSB = 25 μs)
Duration to go in sleep mode. Default value: 0000
SLEEP_DUR[3:0]
1 LSB = 512 ODR
1. Configuration of this bit affects TIMESTAMP0_REG (40h), TIMESTAMP1_REG (41h),
TIMESTAMP2_REG (42h), STEP_TIMESTAMP_L (49h), STEP_TIMESTAMP_H (4Ah), and
STEP_COUNT_DELTA (15h) registers.
82/104
DocID028370 Rev 4