E3A-MA3A-H11 Datasheet, PDF (24/48 Pages) Magnetrol International, Inc. – Digital E3 Modulevel with HART® and PACTware™ Communications Liquid Level Displacer Transmitter
2.6.5.3 Measurement Type: Density (cont.)
34b
Display
User
Cal Menu
35
AdjSnrLo
36
AdjSnrHi
37
Conv Fct
xxxx
38
Scl Ofst
xxx
39
LVDT%
xx.xx %
40
Chan 0
41
Chan 1
42
NSP Value
43
ElecTemp
xxx F
44
Max Temp
xxx F
45
Min Temp
xxx F
Action
Comment
Press to display the User CalSelct = User
Calibration sub-menu
Submenu below
Diagnostic Display
Diagnostic Display
Diagnostic Display
Diagnostic Display
Diagnostic Display
Diagnostic Display
Diagnostic Display
Diagnostic Display
Diagnostic Display
Present temperature in electronics compartment
Diagnostic Display
Maximum electronics temperature recorded
diagnostic Display
Minimum electronics temperature recorded
2.6.5.3 Measurement Type: Density: Factory (display only) or User Calibration sub-menu
Display
1
LVDT%
xx.xx %
2
DrySensr
xx.xx %
3
SnrCalLo
xx.xx %
4
SGCalLo
xx.xx sg
5
Set 4mA
xx.xx sg
6
SnrCalHi
xx.xx %
7
SGCalHi
xx.xx sg
8
Set 20mA
xx.xx sg
9
Escape
Action
Diagnostic display
Comment
Enter or capture sensor
output for Dry Sensor
Enter or capture sensor
output for Low Cal Point
Enter SpecGrav value
corresponding to SnrCalLo
Press and
sensor output
Press and
sensor output
simultaneously to capture current
simultaneously to capture current
Enter the PV value for 4 mA Specify SG at 4 mA (User Cal menu only)
set point
Enter or capture sensor
output for High Cal Point
Enter SpecGrav value
corresponding to SnrCalHI
Press and
sensor output
simultaneously to capture current
Enter the PV value for 20 mA Specify SG at 20 mA (User Cal menu only)
set point
Press to exit Calibration
sub-menu; Returns to
Factory Menu
24
48-635 E3 Modulevel Displacer Level Transmitter