2.6.5.3 Measurement Type: Density (cont.)
Display
13
Trim 20
xxxx
14
Loop Tst
xx.x mA
15
Capture
4.00 mA
Action
Fine tune 20 mA point
Comment
Adjust setting (Trim 4 value – 4095) to output exactly 20.0 mA
Enter a mA output value
to test loop
Pressing in Data Entry
Mode captures current sg
output as 4.00 mA Set Point
Current SpecGrav set as 4 mA point. Submerge displacer in
lowest SG liquid. Press and to “capture” this SG as your
4 mA density.
In Data Entry Mode, can
(8.00 ≤ loop value ≤ 20.00 mA)
adjust loop value to corre- Set second SpecGrav point between 8.00 mA and 20.00 mA.
16
Capture
20.00 mA
spond to current sg output
for setting 20.00 mA Set
Submerge displacer in higher SG liquid. Press and to
“capture” this SG as your upper loop value between 8.00 and
Point
20.00 mA.
17
New Pass
xxx
Enter new password (0 – 255) Displays encrypted value of present password
Default value = 0
18
Language
(select)
Select from English, Spanish, Language choice for LCD display
French, German
19
E3 ModHT
Ver 1.0
Transmitter Display
Product identification
Firmware version
20
DispFact
(select)
Select “Yes” to display factory Selecting “No” takes you back to beginning of menu
parameter menu as below
Diagnostic Display to view
21
History
(current status)
present status and recent
exceptions
Press to view recent exceptions, up to ten events
Diagnostic Display showing
22
Run Time
xxxx.x h
elapsed time since power on
or History Reset
23
History
Reset
Factory only
24
MeasType
(select)
Factory set
Level, IfcLevel, or Density
25
Model
(select)
Factory set
E3A, E3B, E3C, E3D, E3E, E3F
26
SpringSG
(select)
Factory set
Model specific
27
SprgRate
x.x
Factory set
Model specific
28
SprgMatl
Factory set
Model specific
29
TempLimt
xxx F
Factory set
30
Length
xx.xx lu
Factory set
Model specific—maximum process temperature for which unit is
suitable
Model specific—length of measuring range
31
Diameter
x.xxx in
Factory set
Model specific—outside diameter of displacer
32
Weight
x.xx oz
Factory set
Model specific—weight of displacer
33
CalSelct
(select)
Select Factory or User
Calibration
Selects Calibration Parameters used to calculate the
measured PV
34a
Factory
Cal Menu
Press to display the
CalSelct = Factory
Factory Calibration sub-menu Submenu on page 24
48-635 E3 Modulevel Displacer Level Transmitter