2.6.5.1 Measurement Type: Level Only (cont.)
Display
15
Trim 4
xxxx
16
Trim 20
xxxx
17
Loop Tst
xx.x mA
18
Capture
4.00 mA
Action
Fine tune 4 mA point
Comment
Adjust setting (0 – Trim 20 value) to output exactly 4.0 mA
Fine tune 20 mA point
Adjust setting (Trim 4 value – 4095) to output exactly 20.0 mA
Enter a mA output value
to test loop
Pressing in Data Entry
Mode captures current level
output as 4.00 mA Set Point
Raise liquid level to desired 4 mA point on the displacer. Press
and buttons to “capture” this liquid level as your 4 mA
point.
In Data Entry Mode, can
Raise liquid level to a second level point on the displacer. Press
19
Capture
20.00 mA
adjust loop value to corre-
and buttons to “capture” this liquid level as your upper
spond to current level output loop value between 8.00 and 20.00 mA.
for setting 20.00 mA Set
Point
20
New Pass
xxx
Enter new password (0 – 255) Displays encrypted value of present password
Default value = 0
21
Language
(select)
Select from English, Spanish, Language choice for LCD display
French, German
22
E3 ModHT
Ver 1.0
Transmitter Display
Product identification
Firmware version
23
DispFact
(select)
Select ‘Yes’ to display factory Selecting “No” takes you back to beginning of menu
parameter menu as below
Diagnostic Display to view
24
History
(current status)
present status and recent
exceptions
Press to view recent exceptions, up to ten events
Diagnostic Display showing
25
Run Time
xxxx.x h
elapsed time since power on
or History Reset
26
History
Reset
Factory only
27
MeasType
(select)
Factory set
Level, IfcLevel, or Density
28
Model
(select)
Factory set
E3A, E3B, E3C, E3D, E3E, E3F
29
SpringSG
(select)
Factory set
Model specific—range of SGs for which unit is suitable
30
SprgRate
x.x
Factory set
Model specific
31
SprgMatl
Factory set
Model specific
32
TempLimt
xxx F
Factory set
33
Length
xx.xx lu
Factory set
Model specific—maximum process temperature for which unit is
suitable
Model specific—length of measuring range
34
Diameter
x.xxx in
Factory set
Model specific—outside diameter of displacer
35
Weight
xx.x oz
Factory set
Model specific—weight of displacer
36
CalSelct
(select)
Select Factory or User
Calibration
Selects Calibration Parameters used to calculate the
measured PV.
37a
Factory Cal
Menu
Press to display the
CalSelct = Factory
Factory Calibration sub-menu Submenu on page 19
18
48-635 E3 Modulevel Displacer Level Transmitter