English
Language : 

N2893A Datasheet, PDF (7/29 Pages) List of Unclassifed Manufacturers – InfiniiVision Oscilloscope Probes and Accessories
07 | Keysight | InfiniiVision Oscilloscope Probes and Accessories - Data Sheet
High-voltage Passive
Probes
–  Ideal for measuring up to 30 kV
–  Up to 500 MHz bandwidth
–  100:1 or 1000:1 attenuation
10076C high-voltage probe
The Keysight 10076C 4 kV 100:1 passive
probe gives you the voltage and
bandwidth you need for making
high-voltage measurements. Its compact
design makes it easier to probe today’s
small power electronics components and
its rugged construction means it can
withstand rough handling without
breaking.
Characteristics 10076C
Bandwidth
500 MHz (–3 dB)
Risetime (calculated) < 0.7 ns
Attenuation ratio 100:1
Input resistance
Input capacitance
66.7 MΩ (when
terminated into 1 MΩ)
Approx 3 pF
Maximum input
Compensation
Probe readout
Cable length
4000 Vpk
6 to 20 pF
range
Yes
1.8 m
N2771B high-voltage probe
The N2771B is a 1000:1 divider probe for
the measurement of fast high voltage
signals up to 30 kV dc + peak ac, 10 kV
rms.
The probe’s large size and rugged
construction provide superior protection.
The ground lead is fed through the body of
the probe and protrudes behind the safety
barrier, keeping the ground
connection away from the high voltage.
Typical applications include PMTs, motor
drives, high voltage switches, magnatrons,
and modern projection systems.
Characteristics for N2771B
Ordering information for
Keysight high‑voltage probe
10076C
N2771B
10077A
N2789A
High-voltage probe: includes
one retractable hook tip,
one‑ground‑bayonet, one IC
probing tip, one‑alligator ground
lead, one-spring ground tip and a
compensation screwdriver
High-voltage probe:
includes alligator ground lead,
1‑sharp‑probe tip
Accessory kit for 10076A/C:
includes one retractable
pincher tip, one ground lead, one
insulation cap, two measuring
pins and two colored tags
Spring ground tip for 10076B/C
Bandwidth
Risetime
Attenuation ratio
Input resistance
Input capacitance
Compensation
range
Max. voltage
Operating
temperature
Storage tempera-
ture
Dimensions
50 MHz (–3 dB)
< 7 ns
1000:1
100 MΩ (when terminated
into 1 MΩ)
1 pF
6 to 20 pF
15 kV dc, 10 kV rms,
30 kV dc + peak ac
0 to +50 °C,
80% RH
–20 to +70
2 cm (max width of probe
stem after handle) x 33 cm
7.5 cm (max probe width
at probe handle) x 33 cm
10077A accessory kit for 10076A/B
10076C derating curve
10076C passive probe
N2771B high-voltage probe
N2771B derating curve