These specifications apply to all E5515Es, or E5515C mainframe with Option 003,
and an E1966A test application of firmware revision A.09.13 or higher.
Specifications describe the test set’s warranted performance and are valid over the
entire operation and environmental ranges unless otherwise noted. All specifications
are valid after a 30-minute warm-up period of continuous operation.
Supplemental characteristics are intended to provide additional information useful in
applying the instrument by giving typical, but non-warranted performance parameters.
These characteristics are shown in italics and labeled as typical, or supplemental. All
units shipped from the factory meet these typical numbers at 25 °C ambient without
including measurement uncertainty.
Analog Specifications
CW RF generator
Frequency
Available frequency range
Specified frequency ranges
Accuracy and stability
Test signal
Amplitude
Available output level range
Specified output level range
Absolute output level accuracy
VSWR at RF IN/OUT
Nominal ambient test signal
level accuracy
292 to 2700 MHz
421 to 494 MHz, 800 to 960 MHz, and 1700 to 2000 MHz
Same as timebase reference
CW, AM (56% depth with 20 kHz rate), or DSB-SC
(carrier + upper side-band spaced 20 kHz apart). Requires
approximately 3 seconds to switch between test signal selections
−127 to −10 dBm
−116 to −15 dBm
< ±1.0 dB, typically < ±0.5 dB (Level accuracy at RF generator
output levels > −30 dBm may be degraded by simultaneous
reception and transmission when applied Tx power is > 32 dBm)
< 1.14:1, 400 to 1000 MHz
< ±1.1 dB