33 | Keysight | E1966A 1xEV-DO Terminal Test Application - Technical Overview
Fast Device Tune Measurement
E1999A-202 FDT Enhanced
Fast device tune measurement
Input frequency ranges
Measurement method
411 to 420 MHz
450 to 484 MHz
821 to 934 MHz
1700 to 1980 MHz
Allows user definition of an RF source power output sequence
simultaneously with a Tx power measurement sequence each consisting
of 10 or 20 ms duration steps with user-defined step size. Sequence
can be defined to repeat over a number of frequencies inside of a single
frequency band. This measurement requires a test mode in the mobile
station in order to operate. Measures the total power in a 1.23 MHz
bandwidth centered on the active reverse channel center frequency in
each step period
Measurement data
capture period
0.313 μs
Maximum input level
+37 dBm/1.23 MHz peak (5 W peak)
Measurement range
−61 to +30 dBm, usable to < −69 dBm/1.23 MHz with reduced accuracy
Measurement capture range
Mobile station’s transmit power must be within ±9 dB of the expected
power per the ranging configuration
Measurement accuracy
< ±1 dB 15 to 55 °C for the fast mode, typically < ±0.5 dB. (Calibrated
against average power and within ±10 degrees of calibration tempera-
ture. Calibration must occur between 20 to 55 °C)
Measurement resolution
0.01 dBm/1.23 MHz
Measurement step duration (time) 10 or 20 ms
Number of frequency steps
1 to 20
Number of amplitude steps
1 to 20 steps at each specified frequency
Maximum steps in a sequence Up to 20 out of the possible 40 entries in each table
MS Tx frequency step table
1 to 40 entries, with each value in MHz
MS Tx power step table
1 to 40 entries, with each value in dBm
MS Rx frequency step table
1 to 40 entries, with each value in MHz
MS Rx power step table
1 to 40 entries, with each value in dBm
MS Tx frequency step start index 0 to 39
MX Tx power step start index 0 to 39
MS Rx frequency step start index 0 to 39
MS Rx power step start index 0 to 39
RF generator settling time
< 6.1 ms to be within ±0.1 dB of the final value
RF generator
modulation accuracy
Typically < 3.1%
RF generator level accuracy
Same as listed under 1xEV-DO RF generator specifications
Concurrency capabilities
None
Calibrate function
Calibrates all measurement functions
MS Tx
sequence
20 ms
MS Tx power
sequence
Tx power step
313 μs
~6.6 ms
~13 ms
Ranging
Pre-set interval
Measurement
period
MS Rx
sequence
20 ms
MS Rx power
sequence
Rx power step
6.1 ms
13.9 ms
Settling RSSI measurement interval