English
Language : 

E1976A Datasheet, PDF (25/37 Pages) Keysight Technologies – E1966A 1xEV-DO Terminal Test Application
25 | Keysight | E1966A 1xEV-DO Terminal Test Application - Technical Overview
1xEV-DO RF Analyzer (measurements only) (Continued)
Receiver ranging
Auto mode
Manual mode
Autoranges to the ideal RF power level for the nominally expected open
loop response. Provides calibrated results if actual received power is
within ±9 dB of the expected open loop power
User enters expected power. If the Active mode is selected, the test
set uses closed loop power control to drive the mobile to the expected
power. Otherwise, the mobile’s Tx power must be within ±9 dB of the
expected power to provide calibrated results
1xEV-DO Analyzer
1xEV-DO modulation
Input frequency ranges
Detector type
Maximum input level
Measurement range
Measurement level ranging
Measurement data
capture period
Measurement trigger
Measurement result
Concurrency support
Measurement accuracy
−10 to +30 dBm
Measurement repeatability
Measurement resolution
Zero function
411 to 484 MHz
800 to 1000 MHz
1700 to 2000 MHz
Thermal detector
+37 dBm peak (5 W peak)
−10 to +30 dBm
Auto
10 ms
1.67 ms (slot trigger)
Average power
Average power measurements can be made concurrently with all
1xEV-DO measurements that support concurrency
(Accuracy with 10 internal averages, all up power control bits
and R-TCH rate set to 153.6 kbps):
400 to 500 MHz
< ±6.9%, typically < ±3.0%
800 to 1000 MHz
< ±6.2%, typically < ±3.0%
1700 to 2000 MHz
< ±7.4%, RF out only port < ±8.2%,
typically < ±3.3%
Typically < ±0.05 dB
0.01 dBm
Auto zeroes (no user control)