English
Language : 

4082A Datasheet, PDF (5/21 Pages) Keysight Technologies – Parametric Test System
05 | Keysight | 4082A Parametric Test System - Data Sheet
Switching matrix subsystem
Switching matrix subsystem
Maximum DUT pins
SMU port in test head
Maximum number of instrument ports
Auxiliary ports
Maximum voltage at each port
Maximum current, port to DUT pin
Maximum residual resistance
Extended path:
SMU port in test head
AUX port
Optional HF ports
Extended path
Zero reference
SMU port in test head
GNDU
AUX port
Optional HF ports
Extended path
Through AUX port
Maximum stray capacitance between
DUT pins
Isolation resistance
48 output pins plus one pin for the prober chuck connection (triaxial connector).
Two types of DC switching matrix cards are available: standard low-current and
ultra-low current.
Eight SMUs + one GNDU
- Two ports for low-current measurement (Non-Kelvin)
- Four ports (Kelvin)
- Two ports (Non-Kelvin)
- One port for GNDU (Kelvin)
Six for external instruments (Digital voltmeter, etc.) and two for HS-CMU or
E4980A
- Two triaxial input ports (Force/guard/common, AUX ports 1 and 2)
- Four BNC two-pair input ports (Force/common and sense/common,
AUX ports 3 to 6)
- Two BNC input ports (Force/common, AUX ports 7 and 8, connected to
HSCMU in default)
48 extended paths
The system provides one on/off relay for each path.
±200 V
±200 V (AUX ports 1 and 2)
±100 V (AUX ports 3 to 8)
±100 V (between force and common of each HF port)
±100 V (between two of forces of all HF ports)
±100 V (between any force of HF ports and any force of extended paths)
±100 V (between force and common of each extended path)
±100 V (between any force of the optional HF ports and any force of extended paths)
±200 mV
±1.0 A
±1.6 A
±1.0 A
±0.5 A
±0.5 A
Low current port: Force 1.0 Ω
Kelvin port: Force 1.0 Ω, Sense 2.5 Ω
Non-Kelvin port: Force 1.0 Ω
3 pF (supplemental characteristics)
Low current (with guard): 1× 1015 Ω (supplemental characteristics)