English
Language : 

4082A Datasheet, PDF (19/21 Pages) Keysight Technologies – Parametric Test System
19 | Keysight | 4082A Parametric Test System - Data Sheet
General Speciications
Operating and storage requirements
Temperature range
Operating
5°C to 30°C (no condensation)
Humidity range
Storage
Operating
–20°C to 50°C
–20°C to 60 °C (for an unpacked system)
15% to 70% (no condensation)
Power requirement
Storage
< 80% RH (no condensation)
< 90% RH, < 12 hrs (for an unpacked system)
Nominal line voltage1
200 Vac
208 Vac
220 Vac
240 Vac
Allowable voltage range
180 - 220 Vac
188 - 228 Vac
198 - 242 Vac
216 - 252 Vac
Required maximum current
30 A
24 A
30 A
30 A
Regulatory and standard compliance
EMC
EMC Directive (2004/108/EC)
EN 61326-1
ICES/NMB-001
AS/NZS CISPR 11
Safety
Low Voltage Directive (2006/95/EC)
IEC 61010-1:2001/EN 61010-1:2001
CAN/CSA-C22.2 No. 61010-1-04, C/US
SEMI S2-0703/S8-1103
Certiications
CE, CSA C/US, C-tick, ICES/NMB-001
Dimensions
System cabinet
294 kg (including 3458A, SPGU with 5 x HV-SPGU, system controller)
Test head
166 kg (including 7 MPSMUs, 1 HPSMU, 1 HS-CMU, 48 pins, HF matrix, RF matrix, manipulator
extension shelf and PNA (E8363C) with enclosure, fan, and duct)
Supported probers and recommended probe cards2
Prober
TEL P12XL and Precio
ACCRETECH UF3000 and UF3000EX
Probe card2
JEM (Japan Electronic Material)
MJC (Micronics Japan Co.)
SV Probe
FormFactor
1. Line frequency must be 48 Hz to 63 Hz.
2. Please contact your local sales representative regarding the latest information on recommended probers and probe cards.