SMU coniguration
The default SMU coniguration depends upon the matrix card that is chosen (standard low current or ultra-low current).
Please refer to the tables below, which show the SMU installation coniguration associated with different combinations of
SMU resource options.
SMU installation when using ultra-low current matrix cards
No HPSMU
One HPSMU
Port
number
Installed
SMU
Installation
order
Port
number
Installed
SMU
Installation
order
1
MPSMU
2
1
MPSMU
1
2
MPSMU
Fixed
2
MPSMU
Fixed
3
MPSMU
Fixed
3
HPSMU
Fixed
4
MPSMU
Fixed
4
MPSMU
Fixed
5
MPSMU
1
5
MPSMU
Fixed
6
MPSMU
3
6
MPSMU
2
7
MPSMU
4
7
MPSMU
3
8
MPSMU
5
8
MPSMU
4
SSMMUUinsitnaslltaatilolantwiohnenwuhsienng uulstrian-glowulctruarr-elontwmcautrrixrecnartdms atrix cards
One HRSMU, no HPSMUs
One HRSMU, one HPSMU
Port
number
Installed
SMU
Installation
order
1
MPSMU
2
2
HRSMU
Fixed
3
MPSMU
Fixed
4
MPSMU
Fixed
5
MPSMU
1
6
MPSMU
3
7
MPSMU
4
8
MPSMU
5
Two HRSMUs, no HPSMUs
Port
number
Installed
SMU
Installation
order
1
MPSMU
1
2
HRSMU
Fixed
3
HPSMU
Fixed
4
MPSMU
Fixed
5
MPSMU
Fixed
6
MPSMU
2
7
MPSMU
3
8
MPSMU
4
Two HRSMUs, one HPSMU
Port
number
1
2
3
4
5
6
7
8
Installed
SMU
HRSMU
HRSMU
MPSMU
MPSMU
MPSMU
MPSMU
MPSMU
MPSMU
Installation
order
Fixed
Fixed
Fixed
Fixed
1
2
3
4
Port
number
1
2
3
4
5
6
7
8
Installed
SMU
HRSMU
HRSMU
HPSMU
MPSMU
MPSMU
MPSMU
MPSMU
MPSMU
Installation
order
Fixed
Fixed
Fixed
Fixed
Fixed
1
2
3
Port
number
1
2
3
4
5
6
7
8
Two HPSMUs
Installed
SMU
MPSMU
Installation order
Fixed
MPSMU
Fixed
HPSMU
Fixed
HPSMU
Fixed
MPSMU
Fixed
MPSMU
1
MPSMU
2
MPSMU
3
Port
number
1
2
3
4
5
6
7
8
Port
number
1
2
3
4
5
6
7
8
One HRSMU, two HPSMUs
Installed
SMU
MPSMU
Installation order
Fixed
HRSMU
Fixed
HPSMU
Fixed
HPSMU
Fixed
MPSMU
Fixed
MPSMU
1
MPSMU
2
MPSMU
3
Two HRSMUs, two HPSMUs
Installed
SMU
HRSMU
Installation order
Fixed
HRSMU
Fixed
HPSMU
Fixed
HPSMU
Fixed
MPSMU
Fixed
MPSMU
Fixed
MPSMU
1
MPSMU
2
Note: Installation order indicates the order in which additional MPSMUs must be installed.