Measurement functions
DC current, DC voltage, capacitance and conductance, impedance and differential voltage.
Key measurement features
DC measurements
Measurement unit
Measurement functions
Measurement range
Using two low current SMU ports
Using six standard SMU ports
Capacitance, conductance and impedance measurement
Measurement unit
Measurement functions
HS-CMU
Measurement frequencies
Measurement ranges
DC bias voltage
Measurement functions
Keysight E4980A
Measurement frequencies
Measurement ranges
DC bias voltage
Two terminal differential voltage measurements
Measurement unit
Measurement range
Switching matrix
Number of measurement pins
Instrument ports
HRSMU (High resolution SMU)1
MPSMU (Medium power SMU)
HPSMU (High power SMU)
Spot, sweep, pulse bias, and pulse sweep
1 fA2 to 100 mA, 2 μV to 100 V
10 fA to 1 A3, 2 µV to 200 V3
High speed capacitance measurement unit (HS-CMU)
Keysight E4980A LCR meter
C/G, C/G-V, C/G-V/f, Z/θ and Z/θ-f
1 kHz to 2 MHz, 34 points
1 fF to 100 nF, 0.1 nS to 7.5 mS
±10 V
C/G and C/G-V
1 kHz, 10 kHz, 100 kHz and 1 MHz
1 fF to 100 nF, 0.1 nS to 7.5 mS
±40 V
Keysight 3458A
0.1 μV to 100 V (only when using ultra-low current matrix cards), or 1 μV to
100 V
Between 12 and 48 pins
Note: One additional pin is dedicated for the prober chuck connection.
Up to eight SMUs
One ground unit (GNDU)
Eight auxiliary (AUX) ports (two ports are used for HS-CMU)
48 extended paths
Six optional high-frequency (HF) ports and pulse switch input/output ports
1. Can be used only with ultra-low current matrix cards
2. Using HRSMU. Using MPSMU, 10 fA to 100 mA, 2 μV to 100 V