02 | Keysight | 4082A Parametric Test System - Data Sheet
General Description
Contents
General Description.................................. 2
Specification.............................................. 4
DC Measurement Subsystem SMU......... 9
Capacitance Measurement Subsystem.. 15
Software..................................................... 18
General Specifications.............................. 19
Recommended Conditions for
Ultra-Low Current and Low Voltage
Measurements........................................... 20
The Keysight Technologies, Inc. 4082A parametric test system is designed to perform
fast and precise DC measurements, capacitance measurements, and other high
frequency applications such as ring oscillator measurement. The system supports up
to eight source monitor units (SMUs). Each SMU is self-calibrating, and can be individually
configured to force either current or voltage, as well as simultaneously measure either
current or voltage. The system also supports a fully guarded switching matrix customizable
from 12 to 48 pins. One special additional pin is dedicated as a chuck connection.
The Keysight 4082A can be constructed in either a low-current or an ultra-low current
configuration, depending upon the type of matrix card specified. Only 4082A models
containing the ultra-low current matrix cards can use the high-resolution SMU
(HRSMU).
An optional high-speed capacitance measurement unit (HSCMU) is available for the
4082A, which enables the measurement of capacitance and impedance with unprecedented
speed. External instruments can be integrated into the system via six auxiliary input
ports or forty eight extended path inputs. The extended path inputs allow the user to
connect external signals directly to the DUT pins.
Another option offered for the 4082A is a high-frequency switching matrix. The high-
frequency matrix is organized as two 3 x 24 matrices (six inputs in total), and 1 TO 2
furnished cables may be used on each matrix pair to create one 3 x 48 matrix (three
inputs in total). The system also has one 1.6 A ground unit.