English
Language : 

AMC1210 Datasheet, PDF (27/47 Pages) Burr-Brown (TI) – Quad Digital Filter for 2nd-Order Delta-Sigma Modulator
www.ti.com
Example 1: Register Description Mnemonic
Bit 8
Bit 9
CS1
–
'1'
'0'
W
R
AMC1210
SBAS372A – APRIL 2006 – REVISED OCTOBER 2006
Bit 10
SHS
'0'
RW
The bit position in the
register.
The name of the register
bit. A '–' means Not Used
and therefore a write to
such a bit position will get
lost.
The digit is the reset value.
Indicates if the bit position
is a read-only (R), readable
and writable (RW) or
write-only (W).
REGISTER DESCRIPTIONS
This section describes the functionality of each register and its corresponding bits.
Interrupt Register (address 0x00)
The Interrupt Register contains the 12 interrupt flags together with the acknowledge flags. If an interrupt occurs
(that is, when the output of the comparator filter is above the high level threshold or below the low level
threshold, or when one of the modulators is not functional), the appropriate interrupt flag is set (if enabled). An
interrupt flag is reset when the Interrupt Register is read and the corresponding interrupt source is no longer
active. The acknowledge bits are reset when the corresponding data register is read. Table 13 describes the
Interrupt Register.
Table 13. Interrupt Register
Bit 15
AF4
Bit 14
AF3
Bit 13
AF2
Bit 12
AF1
Bit 11
MF4
Bit 10
MF3
Bit 9
MF2
Bit 8
MF1
Bit 7
IFL4
Bit 6
IFH4
Bit 5
IFL3
Bit 4
IFH3
Bit 3
IFL2
Bit 2
IFH2
Bit 1
IFL1
Bit 0
IFH1
'0'
'0'
'0'
'0'
'0'
'0'
'0'
'0'
'0'
'0'
'0'
'0'
'0'
'0'
'0'
'0'
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
BIT POSITION
15
14
13
12
11
10
9
8
BIT
AF4
AF3
AF2
AF1
MF4
MF3
MF2
MF1
DESCRIPTION
Acknowledge flag for Filter 4.
0: No new data available for Filter 4
1: New data available for Filter 4
Acknowledge flag for Filter 3.
0: No new data available for Filter 3
1: New data available for Filter 3
Acknowledge flag for Filter 2.
0: No new data available for Filter 2
1: New data available for Filter 2
Acknowledge flag for Filter 1.
0: No new data available for Filter 1
1: New data available for Filter 1
Modulator failure flag for Filter 4.
0: Modulator is operating normally for Filter 4
1: Modulator failure for Filter 4
Modulator failure flag for Filter 3.
0: Modulator is operating normally for Filter 3
1: Modulator failure for Filter 3
Modulator failure flag for Filter 2.
0: Modulator is operating normally for Filter 2
1: Modulator failure for Filter 2
Modulator failure flag for Filter 1.
0: Modulator is operating normally for Filter 1
1: Modulator failure for Filter 1
Submit Documentation Feedback