8.3.8.2 Temperature Measurement with a Single RTD
The temperature sensing block can be configured to measure a single RTD by setting the TEMP_MODE bit in
register CONFIG_3 to 1. When the temperature measurement runs in PT1000 mode (TEMP_RTD_SEL = 0), the
first interval corresponds to RREF, the second interval is a redundant measurement on RREF and should be
neglected, and the third interval corresponds to RTD1. This operation is represented in Figure 26.
TRIGGER
Reference
Discard
RTD1
START
STOP
tREF
td1
td2
½ tREF
tRTD1
Figure 26. Temperature Measurement with a Single PT1000
The resistance of RTD1 can be calculated using Equation 3. The time delay between measurements can be
approximated using Equation 4 and Equation 5, with the exception that in this case, td1 is a function of ½ tREF and
td2 is a function of tRTD1.
If the temperature measurement runs in PT500 mode (TEMP_RTD_SEL = 1), the first interval is a redundant
measurement on RREF and should be neglected, the second interval corresponds to RREF, and the third interval
corresponds to RTD1. This operation is represented in Figure 27.
TRIGGER
Discard
Reference
RTD1
START
STOP
td1
td2
2 K tREF
tREF
tRTD1
Figure 27. Temperature Measurement with a Single PT500
The resistance of RTD1 can be calculated using Equation 3. The time delay between measurements can be
approximated using Equation 4 and Equation 5, with the exception that in this case, td1 is a function of tREF and
td2 is a function of tRTD1.
8.4 Device Function Description
8.4.1 Time-of-Flight Measurement Mode
The TOF measurement mode is selected by setting the MEAS_MODE bit in the CONFIG_2 register to 0.
8.4.1.1 Liquid Level or Fluid Identification
The TDC1011 performs a single TOF measurement after receiving a trigger signal and returns to the SLEEP
mode when the measurement is complete.
8.4.2 State Machine
A state machine in the TDC1011 manages the operation of the various measurement modes (see Figure 28). At
power-on, the state machine is reset and most blocks are disabled. After the power-on sequence is complete, the
device goes into SLEEP mode if the EN pin is low or into READY mode if the EN pin is high. In the SLEEP or
READY state, the TDC1011 is able to receive SPI commands to set registers and configure the device for a
measurement mode.