English
Language : 

V-SERIES Datasheet, PDF (43/45 Pages) Keysight Technologies – Infiniium V-Series Oscilloscopes
43 | Keysight | Infiniium V-Series Oscilloscopes - Data Sheet
Infiniium V-Series Performance Characteristics (Continued)
General and environmental
General characteristics
Temperature
Humidity
Altitude
Vibration
Power
Weight
Dimensions
Safety
Pollution degree
Installation category
Measurement category
Environment
Operating: 5 to +40 °C
Non-operating: —40 to +65 °C
Operating: Up to 95% relative humidity (non-condensing) at +40 °C
Non-operating: Up to 90% relative humidity at +65 °C
Operating: Up to 4,000 meters (12,000 feet)
Non-operating: Up to 15,300 meters (50,000 feet)
Operating random vibration 5 to 500 Hz, 10 minutes per axis, 0.21 g(rms)
Non-operating random vibration 5 to 500 Hz, 10 minutes per axis, 2.0 g(rms); resonant search 5 to 500 Hz
Swept sine, 1 octave/minute sweep rate, (0.50 g), 5 minute resonant dwell at 4 resonances per axis
100 to 240 VAC at 50/60 Hz; input power 800 Watts
Frame: 52.2 lbs. (23.7 kg); shipping: 71.7 lbs. (32.5 kg)
Height: 10.5 in (26.6 cm); width: 17.2 in (43.6 cm); depth: 19.4 in (49.2 cm)
IEC 61010-1:2010/EN 61010-1 3rd Edition
CAN / CSA22.2 No. 61010-1-12
UL Std. 61010-1 (3rd Edition)
2
2
1
For indoor use only