15 | Keysight | Infiniium V-Series Oscilloscopes - Data Sheet
Industry’s Most Comprehensive Software (Continued)
Open tightly shut eyes with equalization
Serial data equalization for the Infiniium V-Series provides fast
and accurate equalization using feed-forward equalization (FFE),
continuous time linear equalization (CTLE) and decision feedback
equalization (DFE) modeling in real time. The tool allows you to
use a combination of the equalization, which is found in actual
designs.
Serial data equalization software allows you to input your own
equalizer settings and tap values that are used in your design,
allowing you to simulate the eye opening in your receiver. If you
prefer, the software can find the optimal tap values for you as
well, which you can use as reference for your designs.
Powerful and flexible software trigger
The InfiniiScan software is a powerful tool for identifying signal
integrity issues that hardware triggering is unable to find in your
design. This innovative tool scans through thousands of acquired
waveforms per second to help you isolate signal anomalies,
saving you time and improving your designs. It includes an 80-bit
sequence trigger to find events of interest.
InfiniiScan’s zone touch trigger allows you to draw up to
eight “must intersect” or “must not intersect” zones on the
oscilloscope display to visually determine the event-identify
condition. If you can see the event of interest on the screen,
the zone touch trigger can be used to quickly and easily
isolate it, saving significantly more time than complicated
hardware triggers. Other triggers include non-monotonic edge,
measurement limit search, runt and pulse width.
Serial data equalization opens the closed eye using CTLE and DFE
methods.
Draw up to eight “must intersect” or “must not intersect” zones to trigger
on a specific event and condition.