English
Language : 

804-R010-330 Datasheet, PDF (22/32 Pages) Magnetrol International, Inc. – Kotron® Sentinel II Model 804 RF Level Transmitter
2.5.10 Diagnostics Menu
This menu is used to offer the technician a method of testing the unit’s functionality. It also provides
useful information for troubleshooting purposes.
Follow this step-by-step procedure to use this menu. It is assumed that you are starting from the
RUN mode.
Current Menu Location
Run
Run/Diagnostics
Display
Magnetrol Int’l
x.x level units
Diagnostics
> to select
Run/Diagnostics/
Test Relay
Test Relays
> to select
Run/Diagnostics/
Test Relay
Enter Password
Run/Diagnostics/
Test Relay
Select Relay
~
Action
Normal operation
Access the Diagnostics menu
Test the relay outputs menu
Accept the data entered
Set the relay to test
Run/Diagnostics/
Test 4–20 mA Loop
Test 4–20 mA Loop
> to select
Test the 4–20 mA output
Run/Diagnostics/
Signal Values
Run/Diagnostics/
Signal Values
Run/Diagnostics/
Signal Values
Run/Diagnostics/
Signal Values
Run/Diagnostics/
Signal Values
Run/Diagnostics/
View Version #
Run/Diagnostics
Run
Signal Values
> to select
Access the Signal Values
menu
CCO Int Cap
x.xx uS
Check the function of the
internal oscillator
CCO Open Probe
x.xx uS
Check the connection
between the preamplifier
and the probe
CCO Probe
x.xx uS
Previous Menu
> to select
View Version #
x.xx
Previous Menu
> to select
Magnetrol Int’l
x.x level units
Measure the probe’s
capacitance in the application
View the software revision #
Normal operation
Comments
This is the RUN mode.
Press the arrow keys to scroll down until the
display reads Diagnostics.
This menu allows you to energize/de-energize
each output individually.
Enter a number between 1 and 4. Use the arrow
keys to toggle between Energize and De-energize.
Set the display variable for the RUN mode.
Press ENT twice. Enter the value of the desired
output current. Press ENT. The 4–20 mA output
is now at the desired loop current. Press ENT
when finished.
View actual signal readings from the
electronics, probe, and preamplifier.
This test compares the probe’s capacitance
on the probe and displaying it as a function
of the oscillator.
This test compares the probe’s capacitance on
the probe and displaying it as a function of
the oscillator.
This test is measuring the actual capacitance
on the probe and displaying it as a function of
the oscillator.
Press ENT.
Displays the version number of Sentinel II
software which is installed.
Press DEL two times to return to the RUN mode.
The RUN mode.
19
50-604 Kotron Model 804 Sentinel II RF Level Transmitter