804-R010-330 Datasheet, PDF (18/32 Pages) Magnetrol International, Inc. – Kotron® Sentinel II Model 804 RF Level Transmitter
2.5.7 System Configuration Menu (cont.)
Current Menu Location
Run/System Config/
Probe Calibrate
Run/System Config/
Probe Calibrate
Run/System Config/
Probe Calibrate/
Cal Point #1 (LO)
Display
Probe Calibrate
> to select
Cal Pt. #1 (LO)
> to select
Initialize
> to select
Run/System Config/
Probe Calibrate
Enter Password
Run/System Config/
Probe Calibrate/
Cal Point #1 (LO)
Run/System Config/
Probe Calibrate
Run/System Config/
Probe Calibrate
Run/System Config/
Probe Calibrate/
Cal Point #2 (HI)
Run/System Config/
Probe Calibrate/
Cal Point #2 (HI)
Calibrate
Previous Menu
> to select
Cal Pt. #2 (HI)
> to select
Initialize
> to select
Enter Password
Probe Calibrate
Run/System Config/
Probe Calibrate/
Cal Point #2 (HI)
Run/System Config/
Probe Calibrate
Run System Config
Run
Calibrate
Previous Menu
> to select
Probe Calibrate
> to select
Magnetrol Int’l
x.x level units
Action
Access the Probe calibration
menu
Access the lower calibration
point menu
Reset the Cal Pt. #1 (LO)
to zero
Accept the data entered
Calibrate the lower probe
reference point
Access the upper calibration
point menu
Reset the Cal Pt. #2 (HI)
to zero
Accept the data entered
Calibrate the upper probe
reference point
Normal operation
Comments
This is the start of the PROBE CALIBRATE menu.
The first time the unit is calibrated and when
changing “Cal Pt. #1 (LO)”, press ENT.
Otherwise press the DOWN arrow.
The “Enter Password” display will not appear if
the password has been entered anywhere else
in the general program menu. Use the numeric
keys to enter your password.
Calibrate with the level as low on the probe as
possible. There must be at least 1" of level on
any insulated probe. The level entered may be
the actual level in the vessel or the number of
inches of probe coverage.
Press ENT.
Set up for the upper calibration reference point.
The first time the unit is calibrated and when
changing “Cal Pt. #2 (HI)”, press ENT.
Otherwise press the DOWN arrow.
The “Enter Password” display will not appear if
the password has been entered anywhere else
in the general program menu. Use the numeric
keys to enter your password.
Calibrate with the level as high on the probe as
possible. There must, however, be at minimum
2% (of span) differential between the calibration
points. A greater differential will improve accuracy.
The level entered may be the actual level in the
vessel or the number of inches of probe coverage.
Press ENT.
Press DEL twice to return to the RUN mode.
The RUN mode.
15
50-604 Kotron Model 804 Sentinel II RF Level Transmitter