E1962B Datasheet, PDF (15/26 Pages) Agilent(Hewlett-Packard) – cdma2000/IS-95/AMPS Mobile Test Application
CDMA RF analyzer (measurements only)
Frequency range (reverse channels):
US cellular band
1-799, 991-1023, 1024-1323
US PCS band
1-1199
Korean PCS band
0-599
Japan CDMA band
1-799, 801-1039, 1041-1199,
1201- 1600
IMT-2000 band
0-1199
NMT-450 band
1-300, 539-871, 1039-1473,
1792-2016
Secondary 800 band
0-719, 720-919
US PCS 1.9 GHz band
0-1299
AWS band
0-899
Cellular Upper 700 band
0-240
400MHz European PAMR 1-2016
800MHz PAMR band
0-239
700MHz Public Safety band 0-240
Lower 700MHz band
0-360
Maximum input level: +37 dBm peak (5 W peak)
Input level range: −71 to +30 dBm/1.23 MHz
Receiver ranging:
Auto mode: auto-ranges to the ideal RF power level for the
nominally expected open loop response; provides calibrated
results if actual received power is within ±9 dB of the expected
open loop power
Manual mode: user enters expected power; if the “active”
mode is selected, the test set uses closed loop power control to
drive the mobile to the expected power; otherwise, the mobile’s
Tx power must be within ±9 dB of the expected power to
provide calibrated results
CDMA analyzer
Average power measurement
Input frequency ranges:
411 to 484 MHz
800 to 1000 MHz
1700 to 2000 MHz
Detector types:
Peak detector: in IS-95, R-RC1 and R-RC2 modes
Thermal detector: in R-RC3 and R-RC4 modes
Maximum input level: +37 dBm peak (5 W peak)
Measurement range: −10 to +30 dBm; usable from −10 to
−20 dBm with reduced accuracy (peak detector only)
Measurement level ranging: auto
Measurement data capture period: 10 ms
Measurement result: average power
Concurrency support: average power measurements can be made
concurrently with all CDMA measurements that support
concurrency
Peak detector measurement accuracy (20 to 55 °C, after
calibration, IS-95 or reverse RC1, RC2):
−10 to +30 dBm:
400 to 500 MHz
< ±7.3%, typically < ±3.0%
800 to 1000 MHz
< ±7.0%, typically < ±3.0%
1700 to 2000 MHz
< ±7.4%
RF OUT ONLY port
< ±8.3%, typically < ±4.4%
−10 to −20 dBm:
400 to 500 MHz
typically < ±4.4%
800 to 1000 MHz
typically < ±4.2%
1700 to 2000 MHz
typically < ±4.8%
RF OUT ONLY port is selected
typically < ±5.6%
Thermal detector measurement accuracy (accuracy with 10
internal averages; reverse RC3 or RC4):
−10 to +30 dBm:
400 to 500 MHz
< ±6.6%, typically < ±3.0%
800 to 1000 MHz
< ±6.0%, typically < ±3.0%
1700 to 2000 MHz
< ±7.2%
RF OUT ONLY port
< ±8.2%, typically < ±3.3%
Measurement repeatability: typically < ±0.05 dB
Measurement resolution: 0.01 dBm
Zero function: auto zeroes (no user control)
Tuned channel power measurement
Input frequency ranges:
411 to 420 MHz
450 to 484 MHz
824 to 934 MHz
1700 to 1980 MHz
Measurement method: measures the total power in a 1.23 MHz
bandwidth centered on the active reverse channel center
frequency
Measurement data capture period: 0.3125 ms (very fast mode),
1.25 ms (fast mode), or 10 ms (normal mode)
Measurement trigger: 20 ms clock (frame trigger)
Maximum input level: +37 dBm/1.23 MHz peak (5 W peak)
Measurement range: −61 to +30 dBm, usable to < −69 dBm/1.23
MHz with reduced accuracy
Measurement level ranging: auto and manual
Measurement accuracy: < ±1 dB 15 to 55 °C, typically < ±0.5 dB
for the normal and fast modes
< ±1.1 dB 15 to 55 °C, typically < ±0.5 dB for the very fast mode
(calibrated against average power and within ±10 degrees of
calibration temperature; calibration must occur between 20 to
55 °C)
Measurement resolution: 0.01 dBm/1.23 MHz
Measurement result: channel power in a 1.23 MHz bandwidth
Concurrency capabilities: channel power measurements can be
made concurrently with all CDMA measurements that support
concurrency
Calibrate function: calibrates the channel power measurement
over the entire operating frequency range of the test set against
the average power measurement; no external cabling is
required
Calibration time: typically < 120 s
15