English
Language : 

E3A-JA3A-F11 Datasheet, PDF (50/60 Pages) Magnetrol International, Inc. – Liquid Level Displacer Transmitter
8.6 Specifications
8.6.1 Functional
System Design
Measurement Principle
Input
Measured Variable
Physical Range
User Interface
Keypad
Indication
Digital Communication
Interoperability test kit (ITK Revision)
LAS capable
DEV Revision
Available Blocks
Polarity Sensitive
Damping
Block Execution Time
Power (Measured at instrument terminals)
Fieldbus General Purpose/Explosion Proof
IS/FISCO/FNICO
Housing
Material
Cable Entry
Ingress Protection
Chamber
Materials
Wetted parts
Process connections
Buoyancy – continuous displacement utilizing a precision range spring
Level, determined by LVDT core movement affected by
buoyancy force changes on continuous displacer
Up to 120" (300 cm) based on displacer length (C/F for longer ranges)
3-button menu-driven data entry and system security
2-line × 8-character LCD display
FOUNDATION fieldbus™, H1 (31.25 kbits/sec)
ITK 5.0, LAS capable
Yes, Device type: Basic
0X01
AI_1, PID_1, RB_1, TB_1
No
Adjustable 0-45 seconds
15 msec
9 to 32 VDC (17 mA maximum current draw)
This device provides only Functional Isolation
9 to 32 VDC (17 mA maximum current draw)
This device provides only Functional Isolation
Aluminum A356T6 (<0.20% copper), optional 316 stainless steel
3⁄4" NPT and M20
TYPE 4X, IP66
Carbon steel
316/316L stainless steel
316/316L and Inconel® (spring)
Tank Top:
3", 4", 6" ANSI Flange
Chambered:
11⁄2", 2" NPT
11⁄2", 2" Socketweld
11⁄2", 2" ANSI Flanges
50
48-640 E3 Modulevel Displacer Level Transmitter - FOUNDATION fieldbus™