English
Language : 

N1914A Datasheet, PDF (8/28 Pages) Keysight Technologies – EPM Series Power Meters E-Series and 8480 Series Power Sensors
08 | Keysight | N1913A and N1914A EPM Series Power Meters / E-Series and 8480 Series Power Sensors - Data Sheet
N1913A/14A EPM Series Power Meters Performance Characteristics (continued)
Settling time (continued)
Auto filter, 10 dB decreasing power step for normal and X2 modes (not across the range
switch points for E-Series and N8480 Series sensors).
x2
speed
40 ms
Normal
speed
70 ms
Typical
settling
times
70 ms
210 ms
3.4 s
120 ms
400 s
6.5 s
–40 dBm
–50 dBm
–60 dBm
With E-Series E4412/13A sensors
Maximum
dBm
Sensor
dynamic
range
Minimum
dBm
x2
Normal
speed speed
Maximum
dBm
Typical
settling
times
150 ms
180 ms
400 ms
3.6 s
6.6 s
150 ms
200 ms
1s
6.6 s
13.5 s
20 dBm
10 dBm
10 dBm
10 dBm
5 dB
With N8480 Series sensors
Sensor
dynamic
range
Minimum
dBm
Typical
settling
times
x2
speed
40 ms
120 ms
210 ms
400 ms
40 ms
70 ms
400 ms
3.4 s
6.8 s
Normal
speed
70 ms
210 ms
400 ms
1s
70 ms
120 ms
1s
6.5 s
13 s
+10 dBm
+2 dBm
–4 dBm
-10 dBm
-20 dBm
-30 dBm
-40 dBm
-50 dBm
With E-Series E9300A/01A/04A sensors
Maximum
dBm
High power
path
Sensor
dynamic
range
Low power
path
Minimum
dBm
x2
Normal
speed speed
Typical
settling
times
150 ms
180 ms
350 ms
3.5 s
150 ms
200 ms
500 ms
6.6 s
20 dBm
10 dBm
10 dBm
10 dBm
With 8480 Series sensors
Maximum
dBm
Sensor
dynamic
range
Minimum
dBm
Typical
settling
times
x2
speed
40 ms
120 ms
210 ms
400 ms
40 ms
70 ms
400 ms
3.4 s
6.8 s
Normal
speed
70 ms
210 ms
400 ms
1s
70 ms
120 ms
1s
6.5 s
13 s
+40 dBm
+3 2 dBm
–26 dBm
-20 dBm
-10 dBm
0 dBm
-10 dBm
-20 dBm
With E-Series E9300B/01B/00H/01H sensor
Maximum
dBm
+20 dBm
+12 dBm
–6 dBm
0 dBm
-10 dBm
–20 dBm
-30 dBm
-40 dBm
High power
path
Sensor
dynamic
range
Low power
path
Minimum
dBm