English
Language : 

N1914A Datasheet, PDF (23/28 Pages) Keysight Technologies – EPM Series Power Meters E-Series and 8480 Series Power Sensors
23 | Keysight | N1913A and N1914A EPM Series Power Meters / E-Series and 8480 Series Power Sensors - Data Sheet
N8480 Series Thermocouple Power Sensor Specifications
The N8480 Series power sensors (excluding Option CFT) measure power levels from
–35 dBm to +44 dBm (316 nW to 25.1 W), at frequencies from 100 kHz to 50 GHz and
have two independent power measurement range (upper and lower range).
Meanwhile, the N8480 sensors with Option CFT only measure power levels from
–30 dBm to +44 dBm (1 μW to 25.1 W) in single range. Similar to the E‑Series power
sensors, the N8480 Series power sensors are also equipped with EEPROM to store
sensor’s characteristics such as model number, serial number, linearity, temperature
compensation, calibration factor, and so forth.
This feature ensures the correct calibration data is applied by any compatible power
meter connected with N8480 Series power sensor, and to ensure the accuracy of the
measurements.
Calibration factor uncertainties
Table 15. N8480 Series calibration factor uncertainty at 25 ºC ± 3 ºC
Frequency
N8481A N8481B N8481H
100 kHz to 10 MHz
–
–
–
10 MHz to 30 MHz
0.82
1.42
0.77
30 MHz to 500 MHz
0.77
1.48
0.89
500 MHz to 1.2 GHz 0.78
1.48
0.89
1.2 GHz to 6 GHz
6 GHz to 14 GHz
0.91
1.58
1.06
1.26
1.77
1.46
14 GHz to 18 GHz
1.59
1.92
1.73
18 GHz to 26.5 GHz
–
–
–
26.5 GHz to 33 GHz
–
–
–
33 GHz to 34 GHz
–
–
–
34 GHz to 35 GHz
–
–
–
35 GHz to 40 GHz
–
–
–
40 GHz to 45 GHz
–
–
–
45 GHz to 50 GHz
–
–
–
N8482A
0.91
0.78
0.77
0.78
0.89
–
–
–
–
–
–
–
–
–
N8482B
1.48
1.43
1.49
1.49
1.56
–
–
–
–
–
–
–
–
–
N8482H
0.89
0.79
0.89
0.89
1.02
–
–
–
–
–
–
–
–
–
N8485A
–
0.82
1.24
1.26
1.35
1.66
1.83
2.67
3.32
–
–
–
–
–
N8487A
–
–
1.33
1.35
1.41
1.61
1.73
2.26
2.58
2.80
2.80
2.80
3.66
4.23
N8486AR
–
–
–
–
–
–
–
–
2.68
3.19
3.19
3.19
–
–
N8486AQ
–
–
–
–
–
–
–
–
–
3.14
3.40
3.14
3.19
3.26