19 | Keysight | N1913A and N1914A EPM Series Power Meters / E-Series and 8480 Series Power Sensors - Data Sheet
E-Series E9300 Average Power Sensor Specifications (continued)
Switch point data
The E9300 power sensors have two paths
as shown in Table 7. The power meter
automatically selects the proper power
level path. To avoid unnecessary switching
when the power level is near the switch
point, switching point hysteresis has been
added.
E9300 “A” suffix sensors example:
Hysteresis causes the low power path
to remain selected until approximately
–9.5 dBm as the power level is increased,
above this power the high power path
will be selected. The high power path
will remain selected until approximately
–10.5 dBm is reached as the signal level
decreases, below this power the low power
path will be selected.
Switching point linearity:
Typically = ± 0.5% (= ± 0.02 dB)
Switching point hysteresis:
0.5 dB typical
Table 11. E9300 Series sensor switch point specification
E9300 sensor suffix Conditions1
Zero set
Zero drift2
Measurement noise3
A
Lower power path (15% to 75% RH)
500 pW
150 pW
700 pW
Lower power path (75% to 95% RH)
500 pW
4,000 pW
700 pW
High power path (15% to 75% RH)
500 nW
150 nW
500 nW
High power path (75% to 95% RH)
500 nW
3000 nW
500 nW
B
Lower power path (15% to 75% RH)
500 nW
150 nW
700 nW
Lower power path (75% to 95% RH)
500 nW
4 μW
700 nW
High power path (15% to 75% RH)
500 μW
150 μW
500 μW
High power path (75% to 95% RH)
500 μW
3000 mW
500 μW
H
Lower power path (15% to 75% RH)
5 nW
1.5 nW
7 nW
Lower power path (75% to 95% RH)
5 nW
40 μW
7 nW
High power path (15% to 75% RH)
5 μW
1.5 μW
5 μW
High power path (75% to 95% RH)
5 μW
30 mW
5 μW
1.  RH is the abbreviation for relative humidity.
2.  Within 1 hour after zero set, at a constant temperature, after a 24-hour warm-up of the power meter with power sensor connected.
3.  The number of averages at 16 for normal mode and 32 for x2 mode, at a constant temperature, measured over a one minute interval and
two standard deviations.