N1914A Datasheet, PDF (7/28 Pages) Keysight Technologies – EPM Series Power Meters E-Series and 8480 Series Power Sensors
07 | Keysight | N1913A and N1914A EPM Series Power Meters / E-Series and 8480 Series Power Sensors - Data Sheet
N1913A/14A EPM Series Power Meters Performance Characteristics (continued)
Settling time1
Manual filter, 10-dB decreasing power step for normal and x2 modes (not across range
switch points for E-Series and N8480 Series sensors).
Table 3. Settling time
Number of averages
1
2
4
8
16
32
64
128
256
512
1024
Settling time with E-Series sensors (s)
Normal mode
x2 mode
0.08 0.13 0.24 0.45 1.1
1.9
3.5
6.7
14
27
57
0.07 0.09 0.15 0.24 0.45 1.1
1.9
3.6
6.7
14
27
Settling time with N8480 Series sensors (s)
Normal mode
x2 mode
0.15 0.2
0.3
0.5
1.1
1.9
3.4
6.6
13
27
57
0.15 0.18 0.22 0.35 0.55 1.1
1.9
3.5
6.9
14.5 33
Settling time with 8480 Series sensors (s)
Normal mode
x2 mode
0.15 0.2
0.3
0.5
1.1
1.9
3.4
6.6
13
27
57
0.15 0.18 0.22 0.35 0.55 1.1
1.9
3.5
6.9
14.5 33
E-Series sensors In FAST mode (using free run trigger), within the range –50 dBm to +17 dBm, for a 10 dB decreasing power step, the
settling time is:
N1913A: 10 ms2
N1914A: 20 ms2
1.  Settling time: 0 to 99% settled readings over the GPIB.
2.  When a power step crosses through the sensor’s auto-range switch point, add 25 ms. Refer to the relevant sensor manual for switch point information.