N1914A Datasheet, PDF (17/28 Pages) Keysight Technologies – EPM Series Power Meters E-Series and 8480 Series Power Sensors | |||
|
◁ |
17 | Keysight | N1913A and N1914A EPM Series Power Meters / E-Series and 8480 Series Power Sensors - Data Sheet
E-Series E9300 Average Power Sensor Specifications (continued)
Power linearity*
Table 9. E9300 Series power linearity (after zero and cal at ambient environmental conditions) sensor
Sensor
Power
Linearity (25 °C ± 10 °C)
E9300A, E9301A, E9304A
â60 to â10 dBm
â10 to 0 dBm
0 to +20 dBm
± 3.0%
± 2.5%
± 2.0%
E9300B, E9301B
â30 to +20 dBm
+20 to +30 dBm
+30 to +44 dBm
± 3.5%
± 3.0%
± 2.5%
E9300H, E9301H
â50 to 0 dBm
0 to +10 dBm
+10 to +30 dBm
± 4.0%
± 3.5%
± 3.0%
* After zero and calibration at ambient environmental conditions
Linearity (0 °C to 55 °C)
± 3.5%
± 3.0%
± 2.5%
± 4.0%
± 3.5%
± 3.0%
± 5.0%
± 4.0%
± 3.5%
0.5
0.4
0.3
0.2
0.1
0
â0.1
â0.2
â0.3
â0.4
â0.5
â30 â26 â22 â18 â14 â10 â8 â2 2 6
Power (dBm)
10 14 18 20
1
0.8
0.6
0.4
0.2
0
â0.2
â0.4
â0.6
â0.8
â1
â10 â5 0
5 10 15 20 25 30
Power (dBm)
1
0.8
0.6
0.4
0.2
0
â0.2
â0.4
â0.6
â0.8
â1
â10 â5 0
5 10 15 20 25 30
Power (dBm)
Typical E9300A/01A/04A power linearity at 25 °C, after zero and calibration, with
associated measurement uncertainty
Power range
Measurement uncertainty
â30 to â20 dBm
± 0.9%
â20 to â10 dBm
± 0.8%
â10 to 0 dBm
± 0.65%
0 to +10 dBm
± 0.55%
+10 to +20 dBm
± 0.45%
Typical E9300B/01B power linearity at 25 °C, after zero and calibration, with
associated measurement uncertainty
Power range
Measurement uncertainty
â6 to 0 dBm
± 0.65%
0 to +10 dBm
± 0.55%
+10 to +20 dBm
± 0.45%
+20 to +26 dBm
± 0.31%
Typical E9300H/01H power linearity at 25 °C, after zero and calibration, with
associated measurement uncertainty
Power range
Measurement uncertainty
â26 to â20 dBm
± 0.9%
â20 to â10 dBm
± 0.8%
â10 to 0 dBm
± 0.65%
0 to +10 dBm
± 0.55%
+10 to +20 dBm
± 0.45%
+20 to +26 dBm
± 0.31%
|
▷ |