English
Language : 

N1914A Datasheet, PDF (12/28 Pages) Keysight Technologies – EPM Series Power Meters E-Series and 8480 Series Power Sensors
12 | Keysight | N1913A and N1914A EPM Series Power Meters / E-Series and 8480 Series Power Sensors - Data Sheet
N1913A/14A EPM Series Power Meters Ordering Information (continued)
Options (continued)
Documentation
N191xA-0B1
N191xA-0BF
N191xA-0BW
N191xA-ABA
N191xA-ABJ
Hard copy English language User’s Guide and Installation Guide
Hard copy English language Programming Guide
Hard copy English language Service Guide
Hard copy English language User’s Guide and Programming Guide
Hard copy Japanese localization User’s Guide and Programming Guide
E-Series Power Sensor Specifications
The E-Series of power sensors have their calibration factors stored in EEPROM and
operate over a wide dynamic range. They are designed for use with the EPM Series of
power meters and two classes of sensors are available:
–– CW power sensors (E4412A and E4413A)
–– Average power sensors (E9300 sensors)
E-Series CW Power Sensor Specifications
Widest dynamic range: 100 pW to 100 mW (–70 dBm to +20 dBm)
Table 4. E4410 Series max SWR specification
Model
Maximum SWR
Maximum SWR
Maximum power
E 4412A
E4413A
10 MHz to 18 GHz
50 MHz to 26.5 GHz
*10 MHz to < 30 MHz: 1.22
30 MHz to < 2 GHz: 1.15
2 GHz to < 6 GHz: 1.17
6 GHz to < 11 GHz:1.2
11 GHz to < 18 GHz: 1.27
50 MHz to < 100 MHz: 1.21
100 MHz to < 8 GHz: 1.19
8 GHz to < 18 GHz: 1.21
18 GHz to 26.5 GHz: 1.26
200 mW (+23 dBm)
200 mW (+23 dBm)
* Applies to sensors with serial prefix US 3848 or greater
Connector type
Type-N (m)
APC-3.5 mm (m)