27 | Keysight | N1913A and N1914A EPM Series Power Meters / E-Series and 8480 Series Power Sensors - Data Sheet
U2000 Series USB Power Sensor Specifications (continued)
Maximum SWR
Table 19. U2000 Series USB sensors maximum SWR
Model
Frequency range
U2000A
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 14 GHz
14 GHz to 16 GHz
16 GHz to 18 GHz
Maximum SWR
(25 °C ± 10 °C)
1.15
1.13
1.19
1.22
1.26
U2001A
10 MHz to 30 MHz
1.15
30 MHz to 2 GHz
1.13
2 GHz to 6 GHz
1.19
U2002A
50 MHz to 2 GHz
1.13
2 GHz to 14 GHz
1.19
14 GHz to 16 GHz
1.22
16 GHz to 18 GHz
1.26
18 GHz to 24 GHz
1.30
U2004A
9 kHz to 2 GHz
1.13
2 GHz to 6 GHz
1.19
U2000B
10 MHz to 2 GHz
1.12
2 GHz to 12.4 GHz
1.17
12.4 GHz to 18 GHz
1.24
U2001B
10 MHz to 2 GHz
1.12
2 GHz to 6 GHz
1.17
U2000H
10 MHz to 8 GHz
1.15
8 GHz to 12.4 GHz
1.25
12.4 GHz to 18 GHz
1.28
U2001H
10 MHz to 6 GHz
1.15
U2002H
50 MHz to 8 GHz
1.15
8 GHz to 12.4 GHz
1.25
12.4 GHz to 18 GHz
1.28
18 GHz to 24 GHz
1.30
Maximum SWR
(0 °C to 55 °C)
1.21
1.15
1.20
1.23
1.27
1.21
1.15
1.20
1.15
1.20
1.23
1.27
1.30
1.15
1.20
1.14
1.18
1.25
1.14
1.18
1.17
1.26
1.29
1.17
1.17
1.26
1.29
1.31
For more detailed specifications, refer