English
Language : 

N1055A Datasheet, PDF (7/19 Pages) Keysight Technologies – 35/50 GHz 2/4 Port TDR/TDT For the 86100D DCA-X Series Oscilloscope Mainframe
07 | Keysight | N1055A Remote Head Module 35/50 GHz 2/4 Port TDR/TDT - Data Sheet
TDR Fixture De-Embedding
Standard in FlexDCA software revision A.05.30.xx and newer,
TDR fixture de-embedding enables you to remove the effects of
fixtures, cables or probes connected between the remote heads
and DUT. All you need is the S-parameter file for the device you
want to de-embed. Flex DCA supports both 2-port and 4-port
fixtures, and you can use multiple fixtures on DUTs up to 16 ports
(fixtures cannot be connected together in series).
Using the same fixture, cable or probe on both the input and
output of your DUT? FlexDCA makes it simple to reverse a fixture
connected to any DUT port in the de-embed setup.
With FlexDCA, it is simple to re-assign the port order of
the fixtures to de-embed, making it easy for you to use an
S-parameter file created on any instrument.