English
Language : 

N1055A Datasheet, PDF (15/19 Pages) Keysight Technologies – 35/50 GHz 2/4 Port TDR/TDT For the 86100D DCA-X Series Oscilloscope Mainframe
15 | Keysight | N1055A Remote Head Module 35/50 GHz 2/4 Port TDR/TDT - Data Sheet
N1055A receiver-only operation
The N1055A TDR/TDT module, equipped with 16 bit low-noise
samplers, can also be used as a receiver only and perform
accurate oscilloscope, eye/mask and jitter/noise (86100D-
200/300) measurements. When used with the 86107A or
86100D-PTB precision timebase circuits, the N1055A provides a
high-bandwidth measurement system with < 100 fs rms intrinsic
random jitter.
Easy channel and eye diagram simulations
(86100D Option SIM)
Generate waveform and eye diagram simulations without using
an external pattern generator. 86100D Option SIM provides
integrated embedding and de-embedding channel simulations
using S-parameters generated using the DCA (or imported from
other tools).
The N1055A remote head modules also perform high-performance
oscilloscope, eye/mask, and jitter/noise measurements.
FlexDCA may be used to generate waveform and eye diagram
simulations using S-parameters generated by the TDR/TDT system.
One application, FlexDCA, provides
comprehensive measurement capability
FlexDCA software integrates:
–– Impedance (TDR/TDT)
–– S-parameter (86100D-202)
–– Keysight PLTS automatic fixture removal in TDR/TDR mode
(86100D-BFP)
–– Oscilloscope
–– Eye/mask
–– Jitter/noise (Option 200/300)
–– Equalization (Option 201)
–– Channel simulation (Option SIM)
–– Other powerful features into a single application that runs on
the 86100D mainframe or on a PC (using N1010A FlexDCA)
The N1055A module is supported by all 86100D DCA-X DCA
modes of operation. For more information, refer to:
www.keysight.com/find/flexdca