English
Language : 

N1055A Datasheet, PDF (11/19 Pages) Keysight Technologies – 35/50 GHz 2/4 Port TDR/TDT For the 86100D DCA-X Series Oscilloscope Mainframe
11 | Keysight | N1055A Remote Head Module 35/50 GHz 2/4 Port TDR/TDT - Data Sheet
Fast and Accurate Multi-Port S-Parameters (86100D-202)
Scattering parameters (S-parameters) are generated in real-time within the oscilloscope
for simultaneous display with time domain results. They are mathematically derived
from TDR and TDT measurements using the 86100D Option 202 enhanced impedance
and S-parameter analysis software. The N1055A’s fast rise time enables calibrated
S-parameter measurements to 50 GHz.
Measure and display (in real-time) calibrated S-parameters to 50 GHz.
86100D Option 202 performs single-ended and mixed-mode S-parameter
measurements on up to 16 ports. The software automatically controls the sequencing of
TDR steps on up to 4 ports so a full set of 4-port S-parameters can be generated by the
touch of a button.
Select S-parameter measurements quickly and easily.