English
Language : 

N1055A Datasheet, PDF (16/19 Pages) Keysight Technologies – 35/50 GHz 2/4 Port TDR/TDT For the 86100D DCA-X Series Oscilloscope Mainframe
16 | Keysight | N1055A Remote Head Module 35/50 GHz 2/4 Port TDR/TDT - Data Sheet
Specifications Table
Dual/quad electrical channel modules
Channel options
(Number of channels; F = female; M = male)
Number of channels 2
Electrical channel bandwidth
Transition time
(10 to 90% calculated from TR = 0.35/BW)
Channel-to-channel skew range
RMS noise
Characteristic
Maximum
Scale factor (per division)
Minimum
Maximum
DC accuracy (single marker)
DC offset range (referenced from center of screen)
Input dynamic range (relative to channel offset)
Maximum input signal
Nominal impedance
Electrical input (female or male option) 3
TDR step repetition rate (characteristic) 6
Mainframe timebase
Module timebase – standard
Module timebase – FS1
Sample rate, module time base (characteristic) 6
Standard
FS1
Step rise/fall time (transmitted) 7
Without TDR calibration
With TDR calibration
32F/32M 1, 3
34F/34M 1, 3
52F/52M 1, 3
54F/54M 1, 3
2
4
35 GHz 4, 5
10 ps
2
4
35/50 GHz 5
10/7 ps
150 ps
600 µV
730 µV
600/750 µV
950 µV
1 mV/division
100 mV/division
± 2 mV ± 4% of (reading-channel offset)
± 500 mV
± 400 mV
+2 V/–1 V
50 Ω
2.92 mm
1.85 mm
1 to 250 kHz
1 to 80 kHz
1 to 250 kHz
80 kSa/s
250 kSa/s
< 18 ps
Adjustable from 15 ps, characteristic
< 7 ps
Adjustable from 6 ps, characteristic
1.  Module is supported by 86100D DCA-X mainframe and later.
2.  Upgradable from 2 channel to 4 channel after purchase (return to Keysight).
3.  Connector style is the same on all channels and is selected at time of order.
4.  Upgradable from 35 to 50 GHz after purchase (return to Keysight).
5.  Tuned to be –3 dB (± measurement uncertainty) at stated bandwidth(s) using NIST traceable swept-sine test system.
6.  Flex DCA software auto-selects the mainframe or module timebase based on DUT setup.
7.  Defined as the transition time at the output of the remote head. It is calculated by deconvolving the receiver transition time from the measured transition
time when the remote head is terminated with a short. Measured on a negative TDR step.