English
Language : 

E1993A Datasheet, PDF (33/40 Pages) Keysight Technologies – W-CDMA Mobile Test Application
33 | Keysight | E1963A W-CDMA Mobile Test Application - Technical Overview
Trigger source: Immediate, protocol, RF rise, external, auto
Trigger delay: Settable between ±50 ms
Peak threshold: Settable from –120 to +37 dBm
Peak excursion: Settable from 1.2 to 100 dB
Trace functions: Clear write, max hold, min hold
Detector type: Peak or sample
Tx measurements
Spectrum monitor
Operating modes: Active cell and test mode
Measurement modes: Swept mode or zero span
Frequency ranges: Although the spectrum monitor is available at any frequency supported
by the test set, specifications apply only inside of the calibrated bands: 698 to 1000 MHz,
1400 to 1500 MHz, 1700 to 2000 MHz, and 2480 to 2580 MHz
Frequency spans, resolution bandwidth range:
Span and RBW can be independently set, except for zero span; zero span can only be set
with the RBW combinations shown below
(Specifications only apply for span and RBW combinations shown in the following table):
Span
100 MHz
80 MHz
40 MHz
20 MHz
12 MHz
10 MHz
5 MHz
4 MHz
2.5 MHz
1.25 MHz
500 kHz
125 kHz
0
0
0
RBW
5 MHz
1 MHz
300 kHz
100 kHz
100 kHz
100 kHz
30 kHz
30 kHz
10 kHz
3 kHz
1 kHz
300 Hz
1 MHz
300 kHz
100 kHz
Displayed dynamic range
50 dB
55 dB
60 dB
65 dB
65 dB
65 dB
70 dB
70 dB
75 dB
80 dB
80 dB
80 dB
55 dB
60 dB
65 dB
RBW filter types: Flattop in swept mode, Gaussian in zero span
Zero span sweep time: Settable from 50 μs to 70 ms
Zero span offset time: Settable from 0 to 10 s
Reference level range: Settable from –50 to +37 dBm or automatically determined