English
Language : 

E1993A Datasheet, PDF (27/40 Pages) Keysight Technologies – W-CDMA Mobile Test Application
27 | Keysight | E1963A W-CDMA Mobile Test Application - Technical Overview
Extended range dynamic power measurement
Measurement method: Allows measurement of a UE’s transmitter output power across its
entire dynamic power (up to 90 dB) in one measurement cycle. This measurement requires
the UE be put into a test mode which forces it to transmit up to two power sequences and
analyzes the resulting UE output power using the test set
RF input
Expected power setting =
Maximum power to be
encountered in the
first sequence
Start point
Measurement interval
= ½ timeslot
Step length
RF rise = 1 timeslot
> 20 dB
Re-range
period
= 15 timeslots
Second sequence expected
power setting=
Maximum power to be
encountered in the
second sequence
First
Re-range
sequence period
Second
sequence
Time
Measurement bandwidth: Selectable RRC filter on or off
Measurement range: –61 dBm to +28 dBm/3.84 MHz
Measurement accuracy:
–– ±1.0 dB (typically ±0.5 dB), at top 25 dB of dynamic range
–– ±1.0 dB (typically ±0.55 dB), at top 30 dB of dynamic range
–– ±1.0 dB (typically ±0.55 dB), at top 35 dB of dynamic range with RRC filter on
Measurement trigger: RF rise, external
Temperature range: +20 to +55°C
Temperature drift: Typically < 0.1 dB per 10 °C temperature change