E1993A Datasheet, PDF (13/40 Pages) Keysight Technologies – W-CDMA Mobile Test Application
13 | Keysight | E1963A W-CDMA Mobile Test Application - Technical Overview
EVM measurement accuracy:
–– < 2.8% rms (typically < 2.4% rms) for UE EVM ≥ 1% rms, ≤ 2200 MHz
–– < 3.2% rms (typically < 2.8% rms) for UE EVM ≥ 1% rms, 2300 to 2580 MHz
Measurement triggers: Auto, protocol, immediate, external, and HS-DPCCH
HS-DPCCH trigger alignment: Adjustable over
–– Subframes: 0 to 5
–– Timeslots: Ack, Nack or CQI
–– Subslots: 0 to 0.5 timeslots
Other reported parameters with EVM:
–– Frequency error
–– Magnitude error
–– Phase error
–– Origin offset
–– Timing error
–– Peak code domain error
Frequency error measurement range: ±1 kHz
Residual frequency error:
< ±(5 Hz + timebase accuracy) for a measurement interval of 1 timeslot
< ±(7 Hz + timebase accuracy) for a measurement interval of 0.5 timeslot
Frequency error measurement accuracy:
505
504
503
Typical UE
EVM range
502
501
500
0
2
4
6
8 10 12 14 16
UE EVM (% rms)
Peak code domain error accuracy: < ±0.4 dB for code power levels > –25 dB
Timing error measurement range: ±10 μs
Timing error measurement accuracy: < ±0.5 chips (±130 ns)
http://wireless.keysight.com/rfcomms/refdocs/wcdma/wcdma_meas_wfrmqual_desc.html_CIHBBHDJ
IQ tuning
All measurements found in the waveform quality measurement are also available in the IQ
tuning measurement; the specifications are the same in both measurements.
http://wireless.keysight.com/rfcomms/refdocs/wcdma/wcdma_meas_iqtuning_desc.html