37 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
Dual-channel WaveGen — built-in function/arbitrary waveform generator (Specification are typical) (Continued)
Amplitude
Range: Minimum
20 mVpp if |offset| ≤ 0.5 Vpp into Hi-Z 1
10 mVpp if |offset| ≤ 0.5 Vpp into 50 Ω 1
Range: Maximum
10 Vpp except, 9 Vpp if sinc or cardiac, 7.5 Vpp if Gaussian pulse into Hi-Z;
5 Vpp/4.5 Vpp into 50 Ω
Resolution
100 μV or 3 digits, whichever is higher
Accuracy
1.5% (frequency = 1 kHz)
DC offset
Range
± 5 V into Hi-Z, except ± 4 V if sine wave, ± 2.5 V if sinc, cardiac, or Gaussian
pulse into Hi-Z
± 2.5 V into Hi-Z, except ± 2 V if sine wave, ± 1.25 V if sinc, cardiac, or
Gaussian pulse into 50 Ω
Resolution
Larger of 250 uV or 3 digits
Accuracy (waveform modes)
± 1.5% of offset setting ± 1% of amplitude ± 1 mV
Accuracy (DC mode)
± 1.5% of offset setting ± 3 mV
Main output
Impedance
50 Ω typical
Isolation
Not available, main output BNC is grounded
Protection
Overload automatically disables output
Trigger output
Trigger output available on trig-out BNC
Quick action customization key
Quick measure all
Quick measure statistics reset
Quick mask statistics reset
Quick histogram statistics reset
Quick print
Quick save
Quick email
Quick recall
Quick freeze display
Quick trigger mode
Quick clear display
Displays a popup containing a snapshot of all the single waveform measurements
Resets all measurement statistics and the measurement count
Resets mask statistics and counters
Zeros the histogram counters
Print the current screen image
Saves the current setup, screen image, or data file as specified in the settings menu
Emails the current setup, screen image, or data file as specified in the settings menu
Recalls setup, mask, or reference waveform
Freezes the display without stopping running acquisitions or unfreezes the display if currently frozen.
Waveform intensity preserved
Toggles the trigger mode between auto and normal
Clears the display
Display characteristics
Display
Display mode
Resolution
Graticules
Format
Maximum waveform update rate
Persistence
Intensity gradation
12.1-inch capacitive multi-touch/gesture enabled color TFT LCD
Zone/zoom/annotation mode and waveform placement mode
800 (H) x 600 (V) pixel format (screen area)
8 vertical divisions by 10 horizontal divisions with intensity controls
YT and XY
> 135,000 wfm/s (real time)
> 450,000 wfm/s (real time max update rate)
Off, infinite, variable persistence (100 ms to 60 s)
256 intensity levels
1.  Sinc, cardiac and Gaussian pulse: ± 1.25 V into Hi-Z; ± 625 mV into 50 Ω.