English
Language : 

6000 Datasheet, PDF (31/41 Pages) List of Unclassifed Manufacturers – THE BREAD BOARDING MEDIUM FOR SURFACE MOUNT
31 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
Trigger system
Trigger jitter
Trigger bandwidth
Trigger sensitivity
(internal) 1
Edge
Other modes
1 GHz bandwidth
2.5, 4, and 6 GHz
bandwidth
Trigger sensitivity
(external) 1
Trigger level range
± 1.6 V
±8V
Any channel
External
< 1.0-ps rms with the jitter-free trigger
< 3.0-ps rms without the jitter-free trigger
500 MHz, 1 GHz, 2.5 GHz models: bandwidth of oscilloscope. 4-GHz and 6-GHz models: 3.5 GHz
Bandwidth of oscilloscope or 1 GHz, whichever is smaller
≤ 10 mV/div
DC to 1 GHz
Greater of 1 div or 5 mVpp
> 10 mV/div
DC to 1 GHz
0.6 div
≤ 10 mV/div
DC to 2 GHz
Greater of 1 div or 5 mVpp
2.0 to 3.5 GHz
Greater of 1.5 div or 5 mVpp
> 10 mV/div
DC to 2 GHz
0.6 div
2.0 to 3.5 GHz
1.0 div
40 mVpp DC to 100 MHz, 70 mVpp 100 to 200 MHz
200 mVpp DC to 100 MHz, 350 mVpp 100 to 200 MHz
± 6 div from center screen
8-V range = ± 8 V; 1.6-V range = ± 1.6 V
1.  Denotes warranted specifications; All others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
Trigger type selections
InfiniiScan Zone (hardware zone qualifier)
Edge
Edge then edge (B trigger)
Pulse width
Pattern
Or
Trigger on user-defined zones drawn on the display. Applies to one analog channel at a time.
Specify zones as either “must intersect” or “must not intersect.” Up to two zones. > 160,000 wfm/
sec update rate
Supported modes: Normal, peak detect, high resolution, max update rate
Also works simultaneously with the serial decodes and mask/limit test
Trigger on a rising and falling edge of any source, alternating or either edge of analog and digital
channels
Arm on a selected edge, wait a specified time, then trigger on a specified count of another selected
edge. Minimum 4 ns
Trigger on a pulse on a selected channel, whose time duration is less than a value, greater than a
value, or inside a time range
–– Minimum duration setting: 2 ns
–– Maximum duration setting: 10 s
–– Range minimum: 10 ns
Trigger when a specified pattern of high, low, and don’t-care levels on any combination of analog,
digital, or trigger channels is [entered | exited]. Pattern must have stabilized for a minimum of 2 ns to
qualify as a valid trigger condition
–– Minimum duration setting: 2 ns
–– Maximum duration setting: 10 s
Trigger on any selected edges from available sources (analog and digital channels only).
Bandwidth is 500 MHz