English
Language : 

6000 Datasheet, PDF (28/41 Pages) List of Unclassifed Manufacturers – THE BREAD BOARDING MEDIUM FOR SURFACE MOUNT
28 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet
Performance Characteristics
DSOX/MSO 6000 X-Series digital storage/mixed signal oscilloscopes
6000 X-Series specification overview
Half channel bandwidth 1 (–3 dB)
Full channel bandwidth 1 (–3 dB)
Full channel equivalent time bandwidth 1 (–3 dB)
Calculated rise time (10 to 90%)
Input channels
DSOX6002A
DSOX6004A
MSOX6002A
MSOX6004A
Maximum sample rate
Maximum memory depth
Display size and type
Waveform update rate
Typical noise floor at 1 mV/div, 50 Ω
1 GHz
2.5 GHz
4 GHz
1 GHz
2.5 GHz
4 GHz
N/A
N/A
N/A
≤ 350 ps
≤ 140 ps
≤ 112.5 ps
2
4
2 + 16
4 + 16
20 GSa/s half channels, 10 GSa/s full channels
Standard 4 Mpts, Standard segment memory
12.1-inch capacitive multi-touch/gesture-enabled display
> 450,000 waveforms per second
115 μVrms
150 μVrms
150 μVrms
6 GHz
4 GHz
6 GHz
≤ 75 ps
210 μVrms
Vertical system analog channels
Hardware bandwidth 1 MΩ
limits
50 Ω
Input coupling
Input impedance
Input sensitivity range 1 MΩ
50 Ω
Vertical resolution
Maximum input voltage 1 MΩ
50 Ω
DC vertical gain accuracy 1
DC vertical offset accuracy
Channel-to-channel isolation
Offset range
1 MΩ
50 Ω
Dynamic range
1 MΩ
Noise floor at 50 Ω
ESD tolerance
50 Ω
1 mV/div
10 mV/div
100 mV/div
1 V/div
20 MHz, 200 MHz (selectable per channel)
20 MHz, 200 MHz, 1.5 GHz, 3 GHz (selectable per channel)
AC, DC
Selectable: 1 MΩ ± 1% (14 pF), 50 Ω ± 3%
1 mV/div to 5 V/div 2 (200 MHz bandwidth limit at ≤ 2 mV/div)
1 mV/div to 1 V/div 2
8 bits (measurement resolution is 12 bits with averaging)
300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
Frequency derating (assumes sine wave input): 400 Vpk until 40 kHz
Then derates at 20 db/dec until 6 Vpk
50 Ω: ± 5Vpk max
± 3 div: 1.5% of full scale (typical)
± 4 div: 2.5% of full scale (warranted) 2
± 0.1 div ± 2 mV ± 1% of offset setting (valid for an offset of ≤ ± 12 divisions)
≥ 100:1 (DC to 1 GHz), ≥ 30:1 (> 1 GHz)
± 5 V (1 to < 10 mV/div), ± 20 V (10 to ≤ 200 mV/div), ± 100 V (> 200 mV/div)
± 12 div or ± 0.8 V, whichever is smallest (≤ 100 mV/div)
± 12 div or ± 4 V, whichever is smallest (> 100 mV/div)
± 8 divisions from center screen (≤ 100 mV/div), 2nd harmonic distortion of - 40 dbc
± 4 divisions from center screen (> 101 mV/div), 2nd harmonic distortion of - 23 dbc
(For a 10:1 probe on the 1 MΩ input, vertical scaling is multiplied by 10)
± 8 divisions from center screen
1 GHz
2.5 GHz
4 GHz
6 GHz
115 µVrms
150 µVrms
150 µVrms
210 µVrms
330 µVrms
355 µVrms
350 µVrms
395 µVrms
3.15 mVrms
3.25 mVrms
3.20 mVrms
3.35 mVrms
31.5 mVrms
32.5 mVrms
32 mVrms
33.5 mVrms
± 2 kV (on input BNCs)
1.  Denotes warranted specifications; All others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
2.  1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.