13 | Keysight | LTE and LTE-Advanced FDD/TDD X-Series Measurement Application - Technical Overview
Key Speciications (continued)
Description
PXA
Channel power
Minimum power at RF input
Power accuracy
Power accuracy (95% conidence)
Measurement loor (@ 10 MHz BW)
± 0.63 dB
± 0.19 dB
–81.7 dBm (nom)
Transmit on/off power (only applies to N9082B/W9082B)
Burst type
Measurement time
Dynamic range for 5 MHz BW 1
124.5 dB (nom)
Adjacent channel power
Minimum power at RF input
Accuracy
Radio
Offset
frequency
MS
Adjacent
± 0.07 dB (5 MHz)
± 0.11 dB (10 MHz)
± 0.21 dB (20 MHz)
BTS
Adjacent
± 0.23 dB (5 MHz)
± 0.33 dB (10 MHz)
± 0.52 dB (20 MHz)
BTS
Alternate
± 0.11 dB (5 MHz)
± 0.21 dB (10 MHz)
± 0.40 dB (20 MHz)
Dynamic range E-UTRA
Offset
Adjacent
Adjacent
Adjacent
Alternate
Alternate
Alternate
Dynamic range UTRA
Offset
2.5 MHz
2.5 MHz
2.5 MHz
Channel BW
5 MHz
10 MHz
20 MHz
5 MHz
10 MHz
20 MHz
83.5 dB (nom)
(Opt ML –8.5 dBm)
82.1 dB (nom)
(Opt ML –8.3 dBm)
Not available
86.7 dB (nom)
(Opt ML –8.5 dBm)
83.7 dB (nom)
(Opt ML –8.3 dBm)
Not available
Channel BW
5 MHz
10 MHz
20 MHz
86.2 dB (nom)
(Opt ML –8.5 dBm)
84.2 dB (nom)
(Opt ML –8.3 dBm)
Not available
7.5 MHz
7.5 MHz
7.5 MHz
5 MHz
10 MHz
20 MHz
87.3 dB (nom)
(Opt ML –8.7 dBm)
87.0 dB (nom)
(Opt ML –8.4 dBm)
Not available
MXA
EXA
–50 dBm (nom)
± 0.82 dB
± 1.04 dB
± 0.23 dB
± 0.27 dB
–79.7 dBm (nom)
–76.7 dBm (nom)
Trafic, UpPTS, DwPTS, SRS, PRACH
Up to 20 slots
124.5 dB (nom)
122.5 dB (nom)
—36 dBm (nom)
± 0.13 dB (5 MHz)
± 0.20 dB (10 MHz)
± 0.38 dB (20 MHz)
± 0.16 dB (5 MHz)
± 0.24 dB (10 MHz)
± 0.41 dB (20 MHz)
(ACPR range –33 to –27 dBc with Opt ML)
± 0.57 dB (5 MHz)
± 0.82 dB (10 MHz)
± 1.19 dB (20 MHz)
± 1.03 dB (5 MHz)
± 1.29 dB (10 MHz)
± 2.04 dB (20 MHz)
(ACPR range –48 to –42 dBc with Opt ML)
± 0.21 dB (5 MHz)
± 0.35 dB (10 MHz)
± 0.65 dB (20 MHz)
± 0.24 dB (5 MHz)
± 0.39 dB (10 MHz)
± 0.74 dB (20 MHz)
(ACPR range –48 to –42 dBc with Opt ML)
74.2 dB (nom)
(Opt ML –18.4 dBm)
73.8 dB (nom)
(Opt ML –18.4 dBm)
71.7 dB (nom)
(Opt ML –18.2 dBm)
77.6 dB (nom)
(Opt ML –18.6 dBm)
75.1 dB (nom)
(Opt ML –18.4 dBm)
72.1 dB (nom)
(Opt ML –18.2 dBm)
70.0 dB (nom)
(Opt ML –16.5 dBm)
69.3 dB (nom)
(Opt ML –16.5 dBm)
68.4 dB (nom)
(Opt ML –16.3 dBm)
75.8 dB (nom)
(Opt ML –16.6 dBm)
73.2 dB (nom)
(Opt ML –16.3 dBm)
70.3 dB (nom)
(Opt ML –16.3 dBm)
75.9 dB (nom)
(Opt ML –18.5 dBm)
76.2 dB (nom)
( Opt ML –18.4 dBm)
75.0 dB (nom)
(Opt ML –18.2 dBm)
78.4 dB (nom)
(Opt ML –18.5 dBm)
78.6 dB (nom)
(Opt ML –18.4 dBm)
78.1 dB (nom)
(Opt ML –18.2 dBm)
70.5 dB (nom)
(Opt ML –16.6 dBm)
70.5 dB (nom)
(Opt ML –16.4 dBm)
71.4 dB (nom)
(Opt ML –16.3 dBm)
76.5 dB (nom)
(Opt ML –16.6 dBm)
76.5 dB (nom)
(Opt ML –16.4 dBm)
75.7 dB (nom)
(Opt ML –16.3 dBm)
CXA
± 1.33 dB
± 0.61 dB
–72.7 dBm (nom)
119.5 dB (nom)
± 0.37 dB (5 MHz)
± 0.63 dB (10 MHz)
± 0.92 dB (20 MHz)
± 2.16 dB (5 MHz)
± 3.03 dB (10 MHz)
± 4.49 dB (20 MHz)
± 0.91 dB (5 MHz)
± 1.55 dB (10 MHz)
± 2.48 dB (20 MHz)
66.8 dB (nom)
(Opt ML –20.3 dBm)
67.6 dB (nom)
(Opt ML –20.3 dBm)
65.0 dB (nom)
(Opt ML –20.3 dBm)
71.1 dB (nom)
(Opt ML –20.3 dBm)
68.0 dB (nom)
(Opt ML –20.3 dBm)
65.0 dB (nom)
(Opt ML –20.3 dBm)
65.8 dB (nom)
(Opt ML –20.3 dBm)
70.6 dB (nom)
(Opt ML –20.3 dBm)
71.1 dB (nom)
(Opt ML –20.3 dBm)
71.1 dB (nom)
(Opt ML –20.3 dBm)
71.9 dB (nom)
(Opt ML –20.3 dBm)
71.8 dB (nom)
(Opt ML –20.3 dBm)
1. This dynamic range is for the case of 5 MHz information bandwidth. For other information bandwidths, the dynamic range can be derived using the following
equation: Dynamic Range = Dynamic Range for 5 MHz – 10*log10 (Info BW/5.0e6).