RLD Configuration with Multiple Devices
Figure 53 shows multiple devices connected to an RLD.
(AVDD+AVSS)
2
Power-Down
Device N
VA1 VA2
(AVDD+AVSS)
2
Device 2
VA1 VA2
www.ti.com
(AVDD+AVSS)
2
Device 1
VA1 VA2
RLDIN/
RLDREF
RLD
OUT
RLDINV
RLDIN/
RLDREF
RLD
OUT
RLDINV
RLDIN/
RLDREF
RLD
OUT
Figure 53. RLD Connection for Multiple Devices
REXT
CEXT
RLDINV
PACE DETECT
The ADS1291 and ADS1292 provide flexibility for PACE detection by using an external hardware. The external
hardware approach is made possible by bringing out the output of the PGA at pins: PGA1P, PGA1N and PGA2P,
PGA2N.
External hardware circuitry can be used to detect the presence of the pulse. The output of the PACE detection
logic can then be fed into the device through one of the GPIO pins. The GPIO data are transmitted through the
SPI port and loaded 2 tCLKs before DRDY goes low.
When in pace detection mode, the chopping ripple can interfere with pace detect in hardware. It is therefore
preffered to chop thee PGA at a higher frequency (32 kHz or 64 kHz). The RC filter at the PGA output,
suppresses this ripple to a reasonable level. Additionally, suppression can be obtained with an additional RC
stage. The trade-off with chopping the PGA at a higher frequency is an increase in the input bias current.
Figure 6 shows bias current versus input voltage for three different chop frequencies.
RESPIRATION
The ADS1292R provides two options for respiration: internal respiration with external clock and internal
respiration with internal clock, as shown in Table 15.
RESP_CTRL
0
1
Table 15. Respiration Control
Internal respiration with internal clock
Internal respiration with external clock
DESCRIPTION
56
Submit Documentation Feedback
Copyright © 2011–2012, Texas Instruments Incorporated
Product Folder Links: ADS1291 ADS1292 ADS1292R