English
Language : 

MDO4000B Datasheet, PDF (16/28 Pages) List of Unclassifed Manufacturers – Mixed Domain Oscilloscopes
Datasheet
Vertical system digital channels
Probe loading
100 kΩ in parallel with 3 pF
Vertical resolution
1 bit
Horizontal system analog channels
Time base range
1 GHz models
≤ 500 MHz models
400 ps to 1000 s
1 ns to 1000 s
Maximum duration at highest
sample rate (all/half channels)
1 GHz models
≤ 500 MHz models
8/4 ms
8/8 ms
Time-base delay time range
-10 divisions to 5000 s
Channel-to-channel deskew range ±125 ns
Time base accuracy
±5 ppm over any ≥1 ms interval
Horizontal system digital channels
Maximum sample rate (Main)
500 MS/s (2 ns resolution)
Maximum record length (Main)
20M points
Maximum sample rate (MagniVu) 16.5 GS/s (60.6 ps resolution)
Maximum record length (MagniVu) 10k points centered around the trigger
Minimum detectable pulse width 1 ns
(typical)
Channel-to-channel skew (typical) 200 ps
Maximum input toggle rate
500 MHz (Maximum frequency sine wave that can accurately be reproduced as a logic square wave. Requires the use of a short
ground extender on each channel. This is the maximum frequency at the minimum swing amplitude. Higher toggle rates can be
achieved with higher amplitudes.)
Trigger system
Trigger modes
Trigger coupling
Trigger holdoff range
Auto, Normal, and Single
DC, AC, HF reject (attenuates >50 kHz), LF reject (attenuates <50 kHz), noise reject (reduces sensitivity)
20 ns to 8 s
16 www.tektronix.com