English
Language : 

MDO4000B Datasheet, PDF (11/28 Pages) List of Unclassifed Manufacturers – Mixed Domain Oscilloscopes
Mixed Domain Oscilloscopes — MDO4000B Series
Specifications
All specifications apply to all models unless noted otherwise.
Model overview
Analog channels
Analog channel bandwidth
Rise time
Sample rate (1 ch)
Sample rate (2 ch)
Sample rate (4 ch)
Record length (1 ch)
Record length (2 ch)
Record length (4 ch)
Digital channels
Spectrum analyzer channels
Spectrum analyzer frequency range
MDO4014B-3
4
100 MHz
3.5 ns
2.5 GS/s
2.5 GS/s
2.5 GS/s
20M
20M
20M
16
1
9 kHz - 3 GHz
MDO4034B-3
4
350 MHz
1 ns
2.5 GS/s
2.5 GS/s
2.5 GS/s
20M
20M
20M
16
1
9 kHz - 3 GHz
MDO4054B-3
4
500 MHz
700 ps
2.5 GS/s
2.5 GS/s
2.5 GS/s
20M
20M
20M
16
1
9 kHz - 3 GHz
MDO4054B-6
4
500 MHz
700 ps
2.5 GS/s
2.5 GS/s
2.5 GS/s
20M
20M
20M
16
1
9 kHz - 6 GHz
MDO4104B-3
4
1 GHz
350 ps
5 GS/s
5 GS/s
2.5 GS/s
20M
20M
20M
16
1
9 kHz - 3 GHz
MDO4104B-6
4
1 GHz
350 ps
5 GS/s
5 GS/s
2.5 GS/s
20M
20M
20M
16
1
9 kHz - 6 GHz
Spectrum analyzer input
Span
Resolution bandwidth range
RBW shape factor (Kaiser)
Reference level
Input vertical range
Vertical position
Vertical units
1 kHz - 3 GHz (MDO4XX4B-3 models) or 1 kHz - 6 GHz (MDO4XX4B-6 models)
Span adjustable in a 1-2-5 sequence
Variable resolution = 1% of the next span setting
The resolution bandwidth range for windowing functions is as follows:
Kaiser (default): 20 Hz - 200 MHz
Rectangular: 10 Hz - 200 MHz
Hamming: 10 Hz - 200 MHz
Hanning: 10 Hz - 200 MHz
Blackman-Harris: 20 Hz - 200 MHz
Flat-top: 30 Hz - 200 MHz
Adjusted in a 1-2-3-5 sequence
60 dB / 3 dB shape factor: ≥ 4:1
Setting range: -140 dBm to +30 dBm in steps of 1 dB
Vertical measurement range: +30 dBm to DANL
Vertical setting of 1 dB/div to 20 dB/div in a 1-2-5 sequence
-100 divs to +100 divs
dBm, dBmV, dBµV, dBµW, dBmA, dBµA
www.tektronix.com 11