English
Language : 

705-5100-F21 Datasheet, PDF (8/52 Pages) Magnetrol International, Inc. – Enhanced Model 705 with PROFIBUS PA™ Digital Output
x
|
{
x
y
2.1.2 Configuration Information
Some key information is needed to configure the
Eclipse transmitter. Complete the following operating
parameters table before beginning configuration.
Display
Probe Model
Question
What probe model is listed on the
model information?
(first four digits of probe model number)
Answer
_____________
Probe Mount Is the probe mounted NPT, BSP,
or flange?
_____________
Measurement What is the desired measurement? Choices
Type
are: Level only, Volume, Interface Level
or Interface Level and Volume.
_____________
Probe Length What probe length is listed on the
model information?
(last three digits of probe model number) _____________
Level Offset The desired level reading when the
liquid is at the end of the probe.
_____________
Dielectric
What is the dielectric constant range
of the process medium? (Lower layer
dielectric for interface applications)
_____________
2.2 QuickStart Mounting
NOTE: Confirm the configuration style and process connection
size/type of the Eclipse transmitter. Ensure it matches the
}
requirements of the installation before continuing with the
QuickStart installation.
x Confirm the model and serial numbers on the nameplates
of the Eclipse probe and transmitter are identical.
~
For applications using the Model 7xS steam probe, it is
mandatory to keep the transmitter and probe matched
as a set.
z
2.2.1 Probe
y Carefully place the probe into the vessel. Align the probe
process connection with the threaded or flanged mounting
on the vessel.
8
57-645 Eclipse Guided Wave Radar Transmitter - PROFIBUS PA™