English
Language : 

705-5100-F21 Datasheet, PDF (30/52 Pages) Magnetrol International, Inc. – Enhanced Model 705 with PROFIBUS PA™ Digital Output
Analog Input block #
AI block channel
measurement type
*AI#mmm*
######uu
Out Scale units
abbreviation
Out Value
Analog Input Out Display
4.5.2.1 AI Out Display Screens
The Analog Input Out values will be conditionally dis-
played as part of the “rotating” home menu screens.
The screens will be formatted as shown where # in the title
is the number of the AI block (1, 2, 3, or 4) and mmm is
one of: “Lvl”, “Vol”, “Ifc”, “IfV”, “---” depending on the
value of the associated AI block’s Channel parameter.
• For example, “AI1Lvl” would be the most commonly used
AI Out screen.
• “AI2---” would be displayed when the channel value is 0
[uninitialized] for AI block 2.
The Out value will be displayed subject to limitations nec-
essary for a 6-character display [999999 > Value > -99999].
Representative examples are shown below:
*AI1Lvl* *AI3Ifc* *AI1Vol*
99.5 cm
0.0 %
999999 L
Local AI Display
AI1 Out
AI2 Out
AI3 Out
AI4 Out
None
AI1
AI2
AI1+AI2
AI3
AI1+AI3
AI2+AI3
AIs1,2,3
AI4
AI1+AI4
AI2+AI4
AIs1,2,4
AI3+AI4
AIs1,3,4
AIs2,3,4
All AIs
Analog Input Out Values
To Be Displayed
Because the Model 705 transmitter has four Analog Input
blocks any or all of which may be used in particular applica-
tions, a Transducer block parameter controls which AI block
Out values will be displayed.
The fieldbus presentation of this parameter will be similar
to that shown at left (host system dependent).
Any or all (or none) of the AI block Out values can be
selected for display on the LCD.
The local LCD version of this parameter is shown different-
ly due to the limitations of the LCD:
LCD label: “AI Disp ”
The default value of the Local AI Display parameter will be
such that AI 1 Out is selected.
30
57-645 Eclipse Guided Wave Radar Transmitter - PROFIBUS PA™