English
Language : 

E5263A Datasheet, PDF (9/15 Pages) Keysight Technologies – 2 Channel IV Analyzer/Source Monitor Unit
09 | Keysight | E5262A 2 Channel IV Analyzer Source Monitor Unit, E5263A 2 Channel IV Analyzer Source Monitor Unit – Data Sheet
Input
An external trigger input signal can be used to do any of the
following:
1. Start a measurement
2. Start a measurement at each sweep step for a staircase
sweep or multi channel sweep measurement
3. Start the source output at each sweep step for a staircase
sweep, pulsed sweep, staircase sweep with pulsed bias, or
multi-channel sweep measurement.
4. Start the pulsed output for a pulsed spot measurement.
5. Recover from a wait state.
Input level: TTL level, negative or positive edge trigger, or TTL
level, negative or positive gate trigger.
Output
An output trigger signal can be sent when one of the following
events occurs:
1. The end of a measurement is reached.
2. The end of a measurement at each sweep step for a staircase
sweep or multi channel sweep measurement is reached.
3. Completion of the source output setup at each sweep step
for a staircase sweep, pulsed sweep, staircase sweep with
pulsed bias, or multi-channel sweep measurement.
4. Completion of the pulsed output setup for a pulsed spot
measurement.
5. A trigger command is issued.
Output level: TTL level, negative or positive edge trigger, or
TTL level, negative or positive gate trigger.
General Purpose Digital I/O
Altitude
Operating: 0 m to 2,000 m (6,561 ft)
Storage: 0 m to 4,600 m (15,092 ft)
Power requirement
AC voltage: 90 V to 264 V
Line frequency: 47 Hz to 63 Hz
Maximum volt-amps (VA)
E5262A: 400 VA
E5263A: 400 VA
Regulatory compliance
EMC:
Safety:
IEC61326-1/EN61326-1
AS/NZS CISPR 11
KC: RRA Notification amending Radio Waves
Act Article 58-2
IEC61010-1/EN61010-1
CAN/CSA-C22.2 No. 61010-1-12, C/US
Certification
CE, CSA, RCM ,KC
Dimensions
E5262A and E5263A: 426 mm W x 150 mm H x 575 mm D
Weight
E5262A: 15 kg
E5263A: 16 kg
16 general-purpose digital I/O signals are available via a 25-pin
DIN connector. These pins can be used as an alternative to the
BNC trigger-in and trigger-out lines to synchronize the E5262A
and E5263A with other instruments. They can also be used as
output and input ports for digital signals. The user can selectively
assign pins to trigger mode or digital I/O mode.
General Specifications
Furnished Accessories
–– USB-GPIB interface (Keysight 82357B)
–– GNDU to Kelvin adaptor (Keysight N1254A-100)
–– Triaxial cables for SMU
–– Triaxial cable for GNDU
–– Interlock cable
–– CD-ROMs (EasyEXPERT install media,
and VXI plug&play driver)
Temperature range
Operating: +5°C to +40°C
Storage: –20°C to +60°C
Humidity range
Operating: 15 % to 80 % RH, non-condensing
Storage: 5 % to 90 % RH, non-condensing
Furnished Software
–– EasyEXPERT group+
See the following section for features and prerequisites.
–– VXI plug&play driver
Supported OS: Windows 7 Professional
(SP1, 32 bit or 64 bit)