E5263A Datasheet, PDF (8/15 Pages) Keysight Technologies – 2 Channel IV Analyzer/Source Monitor Unit
08 | Keysight | E5262A 2 Channel IV Analyzer Source Monitor Unit, E5263A 2 Channel IV Analyzer Source Monitor Unit – Data Sheet
Functions
Front panel operations
Display
–– Display error messages
–– Display spot measurement set value
–– Display spot measurement result
Keypad operations
–– Set GPIB address
–– Set local/remote mode
–– Select measurement channel
–– Set spot measurement set value
–– Start calibration/diagnostics
MPSMU and HPSMU Measurement
Mode Details
Spot measurement mode
Applies voltage or current, then measures voltage or current.
A maximum of two measurement channels can be used.
Pulsed sweep measurement mode
Applies pulsed sweep voltage or current, and measures voltage
or current at each sweep step. An extra sweep source can be
used to force the staircase sweep output synchronized to the
pulsed sweep output.
Staircase sweep with pulsed bias
measurement mode
Applies pulsed voltage or current, and performs staircase sweep
measurement. The staircase sweep output is synchronized to
the pulsed bias. A synchronous staircase sweep source is also
available.
Quasi-pulsed spot measurement mode
Applies quasi-pulsed voltage or current, and measures voltage or
current.
Search measurement mode (liner search
and binary search)
Applies voltage or current, and measures voltage or current.
Repeats this for various output values until the search stop condi-
tion is satisfied. Synchronous output is available.
Staircase sweep measurement mode
Applies staircase sweep voltage or current, and measures voltage
or current at each sweep step. A maximum of two measurement
channels can be used. An extra sweep source forces the staircase
sweep output synchronized to the primary sweep output. Linear
or log sweeps can be performed.
Number of steps: 1 – 1,001
Hold time: 0 – 655.35 s, 1 ms resolution
Delay time: 0 – 65.5350 s, 100 µs resolution
Multi-channel sweep measurement mode
Applies staircase sweep voltage or current using multiple sweep
sources, and perform staircase sweep measurement. Linear or
log sweeps can be performed.
Number of steps: 1– 1,001
Hold time: 0 – 655.35 s, 1 ms resolution
Delay time: 0 – 65.5350 s, 100 µs resolution
Pulsed spot measurement mode
Applies pulsed voltage or current, and measures voltage or current.
Pulse width: 500 µs to 2 s, 100 µs resolution
Pulse period: 5 ms to 5 s 100 µs resolution
–– Period ≥ width + 2 ms (when width ≤ 100 ms)
–– Period ≥ width + 10 ms (when width > 100 ms)
Maximum pulse duty: 50 %
Time Stamp
The E5262A and E5263A support a time stamp function utilizing
an internal quartz clock.
Resolution: 100 µs
Program Memory
The E5262A and E5263A contain (volatile) memory that can be
used to increase test measurement throughput. Program memory
allows the storage of program code in the E5262A and E5263A,
eliminating the need to communicate over the GPIB interface. In
addition, input data can be passed to code sequences stored in
program memory.
Maximum lines of storable code: 40,000
Maximum number of program sequences: 2,000
Output Data Buffer
The number of data points that can be stored in the data buffer
varies with the choice of the output data format.
Minimum number of storable data Points: 34,034
Trigger I/O
Trigger in/out synchronization pulses before and after setting and
measuring dc voltage and current. Arbitrary trigger events can be
masked or activated independently.