04 | Keysight | E5262A 2 Channel IV Analyzer Source Monitor Unit, E5263A 2 Channel IV Analyzer Source Monitor Unit – Data Sheet
Hardware
Specification conditions
The measurement and output accuracy are specified at the
module connector terminals when referenced to the Zero Check
terminal under the following conditions:
1. Temperature: 23°C ± 5°C (double for 5°C to 18°C, and 28°C
to 40°C if not noted otherwise)
2. After 40 minutes warm-up
3. Ambient temperature change less than ± 1°C after auto
calibration execution
4. Measurement made within one hour after auto calibration
execution
5. Averaging (high-speed per-SMU ADC): 128 samples in 1 PLC;
Integration time
6. Filter: ON (for SMUs)
7. Kelvin connection
8. Calibration period: 1 year
Note: This document lists specifications and supplemental infor-
mation for the E5262A and E5263A and its associated modules.
The specifications are the standards against which the E5262A
and E5263A and its associated modules are tested. When the
E5262A and E5263A or any of its associated modules are shipped
from the factory, they meet the specifications. The “supplemen-
tal” information and “typical” entries in the following specifica-
tions are not warranted, but provide useful information about the
functions and performance of the instrument.
E5262A and E5263A Mainframe Specification
The E5262A and E5263A are identical in form and function except for their SMU configuration.
The E5262A supports two high speed medium power SMUs (MPSMUs).
Description
Range of operation
High speed MPSMU
–100 V to 100 V, –200 mA to 200 mA
Minimum resolution
100 µV, 5 pA
The E5263A supports one high speed medium power SMU (MPSMUs) and one high speed high power SMU (HPSMU).
Description
Range of operation
Minimum resolution
High speed HPSMU
High speed MPSMU
–200 V to 200 V, –1 A to 1 A
–100 V to 100 V, –200 mA to 200 mA
100 µV, 5 pA
100 µV, 5 pA
Maximum output power
There are no power restrictions on the E5262A and E5263A main-
frames. Both mainframes support having both of their modules
simultaneously output maximum voltage or current.
Maximum voltage between common and ground
Maximum common to ground voltage must be ± 42 V
Pulse measurement
Pulse width: 500 µsec to 2 s
Pulse period: 5 ms to 5 s
Period ≥ width + 2 ms (when width ≤ 100 ms)
Period ≥ width + 10 ms (when width > 100 ms)
Pulse resolution: 100 µs
Ground unit (GNDU) specification
The GNDU is furnished with the E5262A and E5263A mainframes.
Output voltage: 0 V ± 100 µV
Maximum sink current: 2.2 A
Output terminal/connection: Triaxial connector, Kelvin (remote
sensing)
GNDU supplemental information
Load capacitance: 1 µF
Cable resistance:
For IS ≤ 1.6 A: Force line R < 1 Ω
For 1.6 A < IS ≤ 2.0 A: Force line R < 0.7 Ω
For 2.0 A < IS ≤ 2.2 A: Force line R < 0.35 Ω
For all cases: Sense line R ≤ 10 Ω
Where IS is the current being sunk by the GNDU.