Digital multimeter specifications (continued)
Supplementary specifications
AC voltage
Measurement method
Maximum input voltage
Input impedance
Input protection
Crest factor
Peak input
Response time
Overload ranging
AC coupled true rms
750 Vrms/1200 Vpeak/3 x 107 V-Hz of product
1 MΩ ± 2% range (typical) in parallel with capacitance < 120 pF
750 Vrms on all ranges
For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at
full scale.
300% of range. Limited by maximum input.
Approximately 2.5 s when the displayed reading reaches 99.9% AC rms value of the tested
input signal at the same range.
Will select higher range if peak input overload is detected during auto range. Overload is
reported in manual ranging.
AC current
Measurement method
Maximum input current
Burden voltage and shunt resistance
Input protection
Crest factor
Peak input
Response time
AC coupled true rms
10 mA to 3.0 A DC or AC rms
• < 0.2 V, 10 Ω for 10 mA range
• < 0.2 V, 1 Ω for 100 mA range
• < 0.3 V, 0.1 Ω for 1 A range
• < 0.7 V, 0.01 Ω for 3 A range
Protected with 3.15 A/500 V, FF fuse
For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at
full scale.
300% of range. Limited by maximum input.
Approximately 2.5 s when the displayed reading reaches 99.9% AC rms value of the tested
input signal at the same range.
Resistance
Measurement method
Open circuit voltage
Input protection
Response time
Two-wire, open-circuit voltage limited to < 5 V
< +5.0 Vdc
1000 Vrms on all ranges, < 0.3 A short circuit
Approximately 0.15 seconds for 1 MΩ and ranges below 1 MΩ
Low-resistance
Measurement method
Input protection
Continuity
Measurement method
Open circuit voltage
Audible tone
Input protection
Four-wire; test current is sent from the FORCE terminals and resistance measured at the
SENSE terminals.
• FORCE terminals: 3.15 A/250 V
•
SENSE terminals: 1000 V on all ranges, < 0.3 A short circuit
rms
0.83 mA ± 0.2% constant current source
< +5.0 Vdc
Continuous beeping when reading is less than the threshold resistance of 10 Ω at 1.0 kΩ
range
1000 Vrms on all ranges, < 0.3 A short circuit
8